7
2. CONFIGURING PRIOR TO OPERATION
2.1. Entering setting values for clamp-on PEAs
2.1.1. Open
Settings / Main settings
menu and enter:
-
Transducer: Clamp-on
– PEA type
-
Mount. mode: Z-mode
or
V-mode
– PEA mounting configuration.
Open
Settings / Main settings / Install. settings
menu and enter
the following setting values for each channel:
-
Lcirc
– average outside diameter circumference of the pipeline section
-
Do
– average outside diameter of the pipeline section
-
Dtr
– average outside diameter of the pipeline section in the longitudinal
plane of PEAs
-
Wall mat
– wall material of the pipeline section
-
hw
– average outside diameter of the pipeline section
-
Coat. mat.
– lining material of the pipeline section
-
hcoat
– pipeline section lining thickness
-
Axial dist. calc.
submenu
-
Lx
– measured distance between acoustic centres of the PEA pair
along the pipeline axis
-
d
– value of equivalent asperity of the inner wall of the pipeline section
-
– kinematic viscosity of the liquid in the pipeline
-
U
– phase velocity for the PEA pair.
Axial dist. calc.
submenu is used to enter the ultrasound velocity
in the measured medium
Cx
and read the recommended value
Lx
.
2.2. Defining zero offset dT0
2.2.1. Zero offset
dT0
shall be defined when the flow in the pipeline is fully
stopped. Open
Settings
menu and select
Calibration / Zero calibra-
tion / Automated / Run calibr
line. Start calibration procedure by se-
lecting
Start
value. After calibration is stopped (to do this assign
Stop.
value to
Run calibr
setting) and the device returns to
ZERO CALIBR.
chan. X
window,
dT0
setting will be automatically set to the value cal-
culated by the flowmeter.
In case it is impossible to absolutely stop the flow of liquid in the
pipeline, you may determine zero offset as follows:
- When the flowrate value lies in the range 0.1Qmax
– Qmax the
dT0
set-
ting is defined according to the foregoing procedure
– dT
01
- For clamp-on PEA pairs with connected cables are reciprocally
swapped instead.
The required value of zero offset is calculated from the formula:
dT
0
= 0.5 (
dT
01
–
dT
02
), µs. (1)
2.2.2. The calculated value is assigned to the
dT0
setting in the flowmeter and
recorded to the log. After that, the initial installation of clamp-on PEAs is
restored.
Summary of Contents for UF-5 Series
Page 17: ...17 Fig A 1 Main menu ...
Page 18: ...18 Fig A 2 Menu FLOW DATA ...
Page 20: ...20 Fig A 4 Menu WALL MATER and COAT MATER Fig A 5 Menu Process depend data ...
Page 22: ...22 Fig A 7 Menu PERIPHERY SETTINGS ...
Page 23: ...23 Fig A 8 Menu STATUS LOGS ...
Page 24: ...24 Fig A 9 Menu DATA LOGS ...