Page | 20
Aeroqual AQM60 User Guide
c. A table listing the modules configured will appear. Check the correct sensors in the AQM60
have been configured. These should match the sensors on the invoice and the sensors
listed in the instrument logbook.
If they do not match click “Setup”
”Configuration” and
enter the password “Password”. Move across the relevant modules. Once completed press
“Save” then “Close” (See Section 3.2.2.).
d. Select Setup
Operations.
e. Enter Data Report Rate (2 minutes is the minimum
) and select “Auto Zero Function OFF”
Click save and close to complete. Note that increasing the report time can cause problems
especially during calibration. (see later the section on calibration) It is recommended to leave
the logging frequency at 2 minutes, and to perform averaging at lower frequencies in
software such as Microsoft Excel.
6. Set AQM60 Real Time Clock
f.
Click on Tools
Update Real Time Clock (this will synchronise the AQM60 clock with the
computer date/time)
7. Start data-logging to confirm sensor communication and operation is correct.
g.
Click Data
Table
Real Time to launch real time data table.
h. Click File
Start Data logging. Data will start being displayed in Real Time Table.
2.3.2. System Checks
Controller display correct
With the AQM on, open the door and observe the display on the controller. This should be scrolling
with sensor readings.
TMS setting
Close the door to the AQM and wait for the internal temperature to stabilise.
Check (and record) that the internal temperature is stable and that it has reached the set point
defined in the logbook e.g. Set point = 30
°
C
SD card logging correct
Verify the AQM is data logging correctly to the SD card by clicking “File”
“Download Files” in the
PC software.
A daily log file (DLYYMMDD.AQL) containing the AQM sensor data is created each day.
System events such as power on, configuration updates, calibration events and system faults are
logged to an EVENTLOG.AQL file.
Summary of Contents for AQM60