Page | 11
Aeroqual AQM60 User Guide
Programming Dip
Switch
Programming
Port
1.8. Internal Air Duct
AQM60 units dispatched after July 2013 will include an
internal air duct (chimney). The chimney is a new
feature to improve thermal stability inside the AQM60
base system. Changes to both the heating and cooling
sub-systems ensure optimum airflow around the
enclosure, minimising thermal gradients thus
temperature effects on the sensors. This optimised
airflow has been coupled with enhancements to the
thermal measurement and control sub-system allowing
an even tighter temperature control ban.
The chimney will come pre-installed. However, there will
be times when the chimney may need to be removed
for servicing.
To remove the chimney the two screws, that connect
the chimney to the fan grill on the base of the
enclosure, need to be removed. The chimney should
then slide out and the two connectors behind can be
easily disconnected.
Note 1: The chimney is stuck to the side of the enclosure with double sided tape which may need to
be replaced when removing/replacing the chimney.
Note 2: The NOx/NO2 module can be removed without taking out the chimney by undoing the
mounting plate screws completely and sliding the module out.
1.9. Auxiliary Module (Optional)
The auxiliary module acts as an interface between third party sensors and the AQM60 communication bus.
It is configured with different operating modes which can be selected by using the dipswitches located on the
side of the module. Aeroqual has integrated a number of third party sensors and is able to supply the
auxiliary module preconfigured for your sensors.