MRK-D-0034 V2.0
Aeroqual AQM 65 User Guide
Page | 54
Figure 4-5 To measure the flow rate, remove the inlet filter and attach the inlet flow adaptor.
4.4.3.2
Measure the flow rate of the individual modules
The module flow rates are measured using a high quality flow meter such as the Aeroqual R7 or the
Bios Dry Cal. See Section 4.2
Attach the flow meter to the inlet port of the gas module
Reconnect the inlet fitting after measuring each gas module
Record the flow rate of each module in the Journal by making a manual user entry
Figure 4-6 To measure the individual module flow rates, disconnect the gas inlet and attach the module
inlet to the vacuum port of the R7 flow meter.