Chapter 3 Operation
Connecting the 2201 ProLock
2201 ProLock
19
Use a 50
Ω
shielded RF cable with an N-type connector (male) to
connect to the unit under test; simply screw the connector tight
to the instrument.
If you have a 50
Ω
shielded RF cable with a BNC connector
(male), use an N to BNC adapter to connect the cable to ProLock.
Aeroflex offers an appropriate adapter.
USB
There is a USB plug on the right-hand side of the front panel. It
can be used to connect an external keyboard, a computer mouse,
or a USB flash drive. More USB ports can be found on the rear
panel.
CAUTION
The maximum input power level at the
RF
IN
/
OUT
connector is
3.5 W (35 dBm) continuous or burst level. Higher input levels
may result in serious damage to the instrument.
CAUTION
Only use a 50
Ω
N-type connector to connect to the
RF
IN
port of ProLock. Use of any other connector may result in
damage to the instrument.