This quick start guide has been developed by the NHS and Cabinet Office and should be used in conjunction with the manufacturers Instructions for Use 

CONNECT AND SET UP DEVICE

What controls does the ventilator have? 

(Image 

E

)

3

SET THE CONTROLS ON THE VG70

AEONMED VG70 VENTILATOR: 

QUICK START GUIDE

CONNECT AND SET UP THE VG70

CLASS A ICU

• Check oxygen sensor is 

connected under flap below 

inspiratory port

• Attach HMEF to expiratory port (

D

)

• Connect patient circuit to 

inspiratory port and to HMEF on 

expiratory port. 

• A standard humidification setup is 

possible as shown

• The Y piece is then connected to a 

test lung and then patient.

• The device is largely touch screen menu driven (

image E

with a control/confirm knob to the side of the screen 

adjacent to an alarm mute button.

• The controls are accessed by pressing the main menu 

button (highlighted 

yellow in the image

)

• The standby button/start ventilation is in the bottom right 

hand corner of the screen (highlighted 

blue in the image

)

• These controls are used to input patient data and select the 

standard modes of ICU ventilation and also to adjust the 

parameters, e.g. Tidal volume, Respiratory rate, I:E, Fi02, 

PEEP/CPAP etc, as well as to set limits.

Expiration valve 

lock

Inspiration port

Expiration valve

Expiration port

Oxygen sensor 

storage shell

Nebulizer

• Connect mains power adapter to the power plug
• Connect Oxygen from the wall or cylinder to the 

oxygen input port. 

• Air connection is not required as this ventilator is 

turbine driven

How do I connect the ventilator?

(Image 

A

)

1

How do I set up the breathing system 

and the HME filter?

(Image B)

2

A

Power 

Switch

Oxygen 

input port

B

C

D

E