A310B Fingertip Pulse Oximeter 

 

Operator’s Manual 

 

      

     

            

 

 

 

 

 

File No.: A310B-OPE        V 1.0      20180712     

Summary of Contents for A310B

Page 1: ...A310B Fingertip Pulse Oximeter Operator s Manual File No A310B OPE V 1 0 20180712...

Page 2: ...every 2 hours SpO2 measurements may be adversely affected in the presence of high ambient light Shield the sensor area with a surgical towel or direct sunlight for example if necessary The following...

Page 3: ...to explode CAUTION Keep the operating environment free of dust vibrations corrosive or flammable materials and extremes of temperature and humidity CAUTION Do not operate the unit if it is damp or wet...

Page 4: ...used if the battery cover is closed CAUTION The batteries must be proper disposed according to local regulation after their use CAUTION The batteries must be taken out from the battery compartment if...

Page 5: ...Equipment Batch code Date of manufacture SN Serial NO Information of manufacture including name and address Temperature limitation Bluetooth Indication When the end user wishes to discard this product...

Page 6: ...rmation you should know to protect the equipment from possible damage Note The important information you should know Batch code Date of manufacturer and Serial No are printed on the label on the batte...

Page 7: ...use in internist surgery Anesthesia pediatrics intensive care and etc Oxygen Bar social medical organizations physical care in sports and etc 2 3 Product Features Lightweight for carrying and Easy To...

Page 8: ...measured arterial hemoglobin saturation value SpO2 of the sensor is compared to arterial hemoglobin oxygen SaO2 value determined from blood samples with a laboratory CO oximeter The accuracy of the s...

Page 9: ...invasive saturation of pulse blood oxygen is low accurate Plethysmograph is an instrument for measuring changes in volume within an organ or whole body usually resulting from fluctuations in the amoun...

Page 10: ...em Name Description 1 Power button Turn on the machine 2 OLED Panel Display the SPO2 PR data Plethysmogram etc 3 Battery Compartment 3 2 Display After switch on the OLED display of A310B is as follows...

Page 11: ...the setting through moving the symbol to the back of the Sound Reminder Beep Bluetooth Restore or Brightness Sound Reminder Press the direction button for 1 second move the symbol to the back of Sound...

Page 12: ...User can press the direction button to select the items And press the direction button for 1 second to change the data you need On the Reminder Limit setup menu page figure 3 3 2 when the symbol shown...

Page 13: ...customer can preset the alarm value to the 98 or 99 to check whether it is normal for alarm setting 3 4 Operation 3 4 1 Install battery Installing two AAA batteries into battery cassette in correct po...

Page 14: ...Press power button to turn the Pulse Oximeter on The oximeter will be automatically powered off when no finger in the device for longer than 16 seconds 3 4 3 Read correspondent data from display scre...

Page 15: ...an been regarded as correct if the wave is fluctuated regularly 3 4 5 Date Transmission 1 The current measurement will transmit to the App automatically after data transmission successfully the measur...

Page 16: ...he power and take out the batteries before cleaning Cleaning exterior surface OLED display screen included of the unit with a dry and soft cloth Use 75 density of medical alcohol to clean the surface...

Page 17: ...ambient temperature is 10 50 and humidity is 15 80 Regular inspection to make sure that no obvious damage existed to affect the safety and performance of device No flammable substance overtop or lowe...

Page 18: ...if you can make sure about no problem existing in the product Please go to a hospital timely for exact diagnosis Oxyhemoglobin or heart rate is shown unstably 1 Finger might not be plugged deep enoug...

Page 19: ...batteries 3 Please contact with local customer service center he screen are suddenly off 1 The product is automatically powered off when no signal is detected longer than 8 seconds 2 Power quantity of...

Page 20: ...ernally powered equipment Anti electric Shock Degree Type BF equipment EMC Type B Mode of operation Continuous Operation Enclosure Degree of ingress protection IP22 IP22 means shell of this product ca...

Page 21: ...non condensing Air Pressure 86Kpa 106Kpa Alarm default value Parameter Value Hemoglobin saturation Upper limit 100 bottom limit 94 Pulse rate Upper limit 130 bottom limit 50 Electronics Parameters Par...

Page 22: ...Accuracy Hemoglobin Saturation 3 70 100 Unspecified 70 Pulse rate 1 BPM PI Display 0 20 Resolution 0 1 Measure Accuracy 0 1 0 1 1 20 1 Probe LED Specification Wave Length Radiant Power RED 660 2 nm 1...

Page 23: ...adiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference w...

Page 24: ...may not cause harmful interference 2 this device must accept any interference received including interference that may cause undesired operation 15 21 Information to user Any Changes or modifications...

Page 25: ...ith RF exposure requirements a minimum separation distance of 5mm for body worn must be maintained between the user s body including the antenna Third party belt clips holsters and similar accessories...