GroundFlex™ Adapter Model 6474
17
3.2 Mechanical
Dimensions:
10.7 x 9.76 x 5.12" (272, 248 x 130mm)
Weight:
7 lbs (3.2kg) approx
Case Material:
ULV0 Polypropylene
Terminals:
4mm recessed banana jacks
Case Protection:
EN 60529 - IP53 (cover closed)
Drop Test:
Per EN 61010-1
Vibration Test:
Per EN 61557-1
3.3 Environmental
Operating Temperature:
32° to 113°F (0° to 45°C); 0 to 90% RH
Specified Operating Temperature
(1)
:
0° to 95°F (0° to 35°C); 0 to 75% RH
Storage Temperature:
-40° to 158°F (-40° to 70°C); 0 to 90% RH
Altitude:
< 3000m
(1): This range corresponds to the one defined by standard EN 61557, for which an operating error including the
quantities of influence is defined. When the device is used outside this range, 1.5%/10°C and 1.5% between
75 and 90% RH must be added to the operating error.
3.4 Safety
Electromagnetic Compatibility
This instrument satisfies the EMC and LVD directives required for the CE marking
and product standard IEC 61326-1 (Ed. 97) + A1 (Ed. 98).
Immunity in industrial environment
•
Emissions in residential environment.
•
Electrical safety according to EN 61010-1 (Ed. 2 of 2001)
Measurement according to EN 61557 (Ed. 2 of 2007) parts 1, 4 and 5.
Cat. IV, 50V
Pollution Degree 2
*All specifications are subject to change without notice