background image

8

7 - CAPACITANCE

Never make a capacitance test on a live circuit
Always respect the polarity of electrolytic condensers.
This type of condenser may be sensitive to
temperature, so avoid touching it during measurement.

  

Connect the leads to the multimeter and to the terminals of the

condenser.

  

Place the selector switch on the capacitance meter function 

  

Automatic range selection: read the measured value

Note:

 The bargraph is not operational on capacitance

measurement

  

Freeze the value, by pressing the HOLD button

C.A 5220 only

If necessary, press the following buttons :

  

RANGE to select the ranges manually

  

MIN-MAX to read the min or max values

  

 to light the display

4nF 

(1)

40nF

400nF

4µF

 40µF

Resolution

1pF

10pF

100pF

1nF

10nF

Accuracy

± 3% R

± 3% R

± 5% R

± 15% R

±400cts

±25cts

±10cts

±10cts

Protection

500V rms and 750V peak

(1)  Min measurement: 400 pF

8 - FREQUENCY

(C.A 5220 only)

  

Connect the leads to the multimeter and in parallel to the circuit to

be tested.

  

Place the selector switch on the Hz voltage frequency function.

  

Automatic range selection : read the measured value

Note:

 The bargraph is inhibed when measuring frequency.

   

If necessary, press the following buttons:

- MIN-MAX to read the min or max values

 to light the display

Note : 

extension to 10000cts

Hz

100Hz

1kHz

10kHz

100kHz

400kHz

Resolution

0.01Hz

0.1Hz

1Hz

10Hz

100Hz

Accuracy

± 0.1% of Reading ± 2cts

Permitted

500V rms and 900V peak

overload

Operating

4V to 600V

range (1)

(1)  The RANGE button on this function selects the trigger level: 10mV, 100mV or

1V (10mV is selected when the instrument is switched on).

Summary of Contents for C.A 5210

Page 1: ...italMultimeters Model C A 5210 Model C A 5220 MULTIMETER Hz Velec V OFF mA 10 A M I N MAX RANGE HOLD 10A 400mA FUSED 10 A max FUSED C A 5220 COM 600V CAT III M U L T I M E T E R MAX RANGE n F Hz mVA M...

Page 2: ...Fuse 12 A HBC 600 V min 10 3 x 38 mm Never use on voltage to earth circuits above 600 V with a voltage surge category of over III that is to say fixed industrial and domestic installations ref IEC 664...

Page 3: ...00 counts 40 segment bargraph The range selection is automatic with the possibility of manual selection on the C A 5220 It switches Off automatically if no control or function button is pressed for 30...

Page 4: ...Automatic selection of AC and when it is switched on HOLD To freeze the last measurement digital and bargraph MIN MAX C A 5220 only To successively display the min or max values The C A 5220 then cha...

Page 5: ...DC current ranges via the 10A terminal mA C A 5220 only Current ranges up to 400mAAC DC via the terminal TERMINALS 4mm safety terminals COM common black red 10A C A 5220 only red lead for the 10AAC D...

Page 6: ...mV 1mV 10mV 100mV 1V resolution Bargraph 10mV 100mV 1V 10V 100V resolution V AC Impedance 10M V elec Impedance 270k Accuracy 3 5 R 1 5 of Reading 1ct Permitted 600V rms and 900V peak overload 1 C A 52...

Page 7: ...the 400 range continuous beep for a resistance R 40 6 DIODE TEST Never test a diode on a live circuit Connect the leads to the multimeter and to the terminals of the component to be tested Place the s...

Page 8: ...F 400nF 4 F 40 F Resolution 1pF 10pF 100pF 1nF 10nF Accuracy 3 R 3 R 5 R 15 R 400cts 25cts 10cts 10cts Protection 500V rms and 750V peak 1 Min measurement 400 pF 8 FREQUENCY C A 5220 only Connect the...

Page 9: ...in or max values to light the display DC currents A DC 40mA 400mA 10A 1 Voltage 600mV 4 5V 600mV drop 2 Digital 10 A 100 A 10mA resolution Bargraph 1mA 10mA 1A resolution Accuracy 1 5 of Reading 1ct 1...

Page 10: ...are pressed and for the overload indication Note No beep on button and on switch between V and Velec 10 4 Climatic conditions Temperature use 0 C to 50 C storage 20 C to 60 C Relative humidity use 80...

Page 11: ...10 12A for C A 5220 Cat 2970 21 12 WARRANTY Our guarantee is applicable for three years after the date on which the equipment is made available copy available on request 13 MAINTENANCE For maintenanc...

Page 12: ...tandards For instrument repair and or calibration please call our factory toll free at 800 945 AEMC 800 945 2362 CHAUVINARNOUX Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Tel 800 95...

Reviews: