7 - CAPACITANCE
Never make a capacitance test on a live circuit
Always respect the polarity of electrolytic condensers.
This type of condenser may be sensitive to
temperature, so avoid touching it during measurement.
Connect the leads to the multimeter and to the terminals of the
condenser.
Place the selector switch on the capacitance meter function
Automatic range selection: read the measured value
Note:
The bargraph is not operational on capacitance
measurement
Freeze the value, by pressing the HOLD button
C.A 5220 only
If necessary, press the following buttons :
RANGE to select the ranges manually
MIN-MAX to read the min or max values
to light the display
4nF
(1)
40nF
400nF
4µF
40µF
Resolution
1pF
10pF
100pF
1nF
10nF
Accuracy
± 3% R
± 3% R
± 5% R
± 15% R
±400cts
±25cts
±10cts
±10cts
Protection
500V rms and 750V peak
(1) Min measurement: 400 pF
8 - FREQUENCY
(C.A 5220 only)
Connect the leads to the multimeter and in parallel to the circuit to
be tested.
Place the selector switch on the Hz voltage frequency function.
Automatic range selection : read the measured value
Note:
The bargraph is inhibed when measuring frequency.
If necessary, press the following buttons:
- MIN-MAX to read the min or max values
-
to light the display
Note :
extension to 10000cts
Hz
100Hz
1kHz
10kHz
100kHz
400kHz
Resolution
0.01Hz
0.1Hz
1Hz
10Hz
100Hz
Accuracy
± 0.1% of Reading ± 2cts
Permitted
500V rms and 900V peak
overload
Operating
4V to 600V
range (1)
(1) The RANGE button on this function selects the trigger level: 10mV, 100mV or
1V (10mV is selected when the instrument is switched on).