Digital Flexprobe
®
Models 400D & 4000D
13
CHAPTER 4
SPECIFICATIONS
4.1
Reference Conditions
Quantity of influence
Reference values
Temperature
73.4 ± 5.4°F (
23 ± 3°C
)
Relative humidity
45 to 75% RH
Frequency of the signal measured
40 to 65Hz
Peak factor of the signal measured
√
2
Conductor diameter
≤
5mm
Battery voltage
2.8 to 3.2V
External electric field
none
External DC magnetic field (earth field)
<40 A/m
External AC magnetic field
none
Position of the conductor
centered in the measurement coil
Shape of the measurement coil
nearly circular
4.2 Electrical
Model 400D-6 / 400D-10
Display Range
4A
40A
400A
Measurement Range
0.020 to 3.999A
4.00 to 39.99A
40.0 to 399.9A
Measurement Range (max)
0.100 to 3.999A
4.00 to 39.99A
40.0 to 399.9A
Resolution
1mA
10mA
100mA
Accuracy
±(2% + 10ct)
±(1.5% + 2ct)
±(1.5% + 2ct)
Model 4000D-14
/ 4000D-24
Display Range
40A
400A
4000A
Measurement Range
0.20 to 39.99A
40.0 to 399.9A
400 to 3999A
Measurement Range (max)
1.00 to 39.99A
40.0 to 399.9A
400 to 3999A
Resolution
10mA
100mA
1A
Accuracy
±(2% + 10ct)
±(1.5% + 2ct)
±(1.5% + 2ct)
Summary of Contents for 4000D-14