Data Logger Model DL913 & DL914
Battery Status
lists the charging
current, charging voltage, and
remaining voltage of the battery
§ 3.1
Firmware Revision
lists the latest
firmware revision running on the
instrument.
The instrument’s serial number
will scroll across the bottom of the
screen.
N/A
4.3 RECORDING DATA
The instrument interface allows you to manually start and stop recordings. The
stored data can be uploaded to a computer with DataView for analysis and
report generation.
NOTE:
To set up and run scheduled recordings on the instrument,
consult the Data Logger Control Panel Help.
NOTE:
To successfully complete a 30-day recording on battery
power only, the instrument must be configured for Extended recording
mode. If upon receipt of the instrument the user starts a recording by
pressing
prior to
the instrument being configured for Extended
recording mode, the recording configuration defaults to a 30-day
recording with 1-minute aggregations and Normal recording mode.
Maximum full battery life for Normal recording mode is approximately
4-days. External power would be required to successfully complete
the 30-day recording.
1. Connect the instrument’s three (DL913) or four (DL914) captive current
probes to the distribution system under test.
2. Press
(≥ 3 s) or connect to a external power source to turn the
instrument ON.
3. Check the memory indicator. If it appears full (
), you will need to delete
recordings from memory to create room for the new recording (see § 3.3).
4. Press to display the Start screen, ‘Press Enter’ scrolls across the bottom