112
Power Quality Analyzer PowerPad
®
IV Model 8345
17.8.4 Markings per IEC 62586-1
The marking
PQI-A-PI
means:
PQI-A:
Class A power quality instrument
P:
Portable measuring instrument
I:
Indoor use
17.9 ELECTROMAGNETIC COMPATIBILITY (EMC)
The instrument is in compliance with the requirements of standard IEC/EN
61326-1 or BS EN 61326-1.
■
The instrument is intended for use in an industrial environment
■
The instrument is a Class A product
■
This instrument is not intended for use in residential environments and
may not ensure adequate protection of radio reception in this residential
environment
For AmpFlex
®
and MiniFlex
®
sensors:
■
An (absolute) influence of 2 % may be observed on the current THD
measurement in the presence of a radiated electric field
■
An influence of 0.5 A may be observed on the RMS current measurement in
the presence of conducted radio frequencies
■
An influence of 1 A may be observed on the RMS current measurement in
the presence of a magnetic field
17.10 RADIO EMISSIONS
The instrument complies with directive RED 2014/53/EU and FCC regulations.
The Wi-Fi module is certified in compliance with the FCC regulations under
number XF6-RS9113SB.
17.11 GPL CODE
The source codes are available under GNU GPL (General Public License).