ADWA AD33 User Manual Download Page 1

USER  MANUAL

AD33 • AD34 • AD35 • AD36

Waterproof EC and TDS Testers

w w w . a d w a i n s t r u m e n t s . c o m

INTRODUCTION

AD33AD34AD35 and AD36 are waterproof
EC and TDS testers, that also measure tempera-
ture. The housing has been completely sealed
against humidity.

All EC and TDS readings are automatically
temperature compensated, and temperature val-
ues can be displayed in °C or °F units.

EC (or TDS) calibration is automatic at one
point, while temperature range is factory cali-
brated and can only be adjusted by the user.

The AD33P probe supplied with the meters, is
interchangeable and can be easily replaced by
the user.
The encapsulated temperature sensor allows
fast and accurate temperature measurement and
compensation.
To prolong battery life, these models are pro-
vided with an auto-off feature that turns them
off after 5 minutes of non-use.

Moreover, when batteries become too weak to
ensure reliable readings the meter automatically
turns off.

Each meter is supplied complete with:

• AD33P EC probe
• 4 x 1.5V batteries, button type
• User manual

FRONT PANEL & DISPLAY

1. HOLD indicator
2. Calibration mode indicator
3. Secondary LCD level with measure units
4. Primary LCD level
5. Measurement units for primary LCD level

(µS for AD33-AD35, ppm for AD34-AD36)

IS

T

AD36     07/07

Dear Customer,

Thank you for choosing an Adwa product.

Please read carefully this manual before starting
operations.

These instruments are in compliance with the
EMC Directive 89/336/EEC and Low Voltage
Directive 73/23/EEC for electrical equipments.

For additional technical information, please e-
mail us at [email protected].

1. Dual line LCD
2. ON/OFF button
3. HOLD button
4. EC probe & temperature sensor
5. Probe body
6. Battery compartment (inside)
7. CAL button
8. Clip holder

PROBE & SOLUTIONS

AD33P

Spare EC probe for AD33,
AD34, AD35, AD36

A70031P        1413 µS/cm solution, 20 ml,

25 pcs

A70030P           12.88mS/cm solution, 20 ml

A70032P          1382 ppm (mg/l) solution, 20

ml, 25 pcs

TECHNICAL DATA

Range

0.0 to 60.0°C / 32.0 to 140.0°F
0 to 1999 µS/cm

(AD33)

0 to 1999 ppm

(AD34)

0 to 199.9 µS/cm

(AD35)

0 to 199.9 ppm

(AD36)

Resolution

0.1°C / 0.1°F
1 µS/cm

(AD33)

1 ppm

(AD34)

0.1 µS/cm

(AD35)

0.1 ppm

(AD36)

Accuracy (@20°C/68°F)

±0.5°C / ±1°F
±2% f.s. (EC/TDS)

Calibration

EC & TDS: automatic, 1 point
Temperature: adjustment

Probe

AD33P (included)

Battery Type / Life

4 x 1.5V button type
Approx. 150 hours of use

Auto-off After 5 minutes of non-use

Environment

0 to 50°C (32 to 105°F);
RH 100%

Dimensions/Weight

175.5 x 39 x 23 mm / 100 g

Summary of Contents for AD33

Page 1: ...USER MANUAL AD33 AD34 AD35 AD36 Waterproof EC and TDS Testers w w w a d w a i n s t r u m e n t s c o m ISTAD36 07 07 ...

Page 2: ...emperature sensor allows fast and accurate temperature measurement and compensation To prolong battery life these models are pro vided with an auto off feature that turns them off after 5 minutes of non use Moreover when batteries become too weak to ensure reliable readings the meter automatically turns off Each meter is supplied complete with AD33P EC probe 4 x 1 5V batteries button type User man...

Page 3: ...dary LCD level with measure units 4 Primary LCD level 5 Measurement units for primary LCD level µS for AD33 AD35 ppm for AD34 AD36 1 Dual line LCD 2 ON OFF button 3 HOLD button 4 EC probe temperature sensor 5 Probe body 6 Battery compartment inside 7 CAL button 8 Clip holder ...

Page 4: ... AD33 0 to 1999 ppm AD34 0 to 199 9 µS cm AD35 0 to 199 9 ppm AD36 Resolution 0 1 C 0 1 F 1 µS cm AD33 1 ppm AD34 0 1 µS cm AD35 0 1 ppm AD36 Accuracy 20 C 68 F 0 5 C 1 F 2 f s EC TDS Calibration EC TDS automatic 1 point Temperature adjustment Probe AD33P included Battery Type Life 4 x 1 5V button type Approx 150 hours of use Auto off After 5 minutes of non use Environment 0 to 50 C 32 to 105 F RH...

Page 5: ...Note Before taking any measurement make sure the meter has been calibrated Freeze the display From measurement mode press the HOLD button The H tag lights up and the read ing is frozen on the LCD Press any button to return to normal mode Change the temperature unit C F To change the temperature measure unit press the HOLD button for 2 seconds Turn the meter off From measurement mode press the ON O...

Page 6: ...ion solution The CAL tag lights up to indicate that the meter is in calibration mode Note The calibration solution value must be within the meter measurement range Wait a few seconds for the reading to stabi lize then use the CAL and HOLD buttons to set the solution value CAL to increase the value HOLD to decrease it The calibration will be automatically per formed To confirm the calibration press...

Page 7: ...erature reading Release the buttons and immerse the probe and an accurate reference thermometer into the sample solution Read the temperature measured by the refer ence thermometer Wait a few seconds for the reading to stabi lize on the tester then adjust it accordingly with the thermometer using the CAL and HOLD buttons CAL to increase the value HOLD to decrease it To confirm press ON OFF for 2 s...

Page 8: ...ge appears on the LCD then the meter turns off Batteries must be replaced Unscrew and release the probe body Take out the battery compartment and carefully replace all four batteries while pay ing attention to their po larity Reattach and tighten the probe body properly to ensure a watertight seal ...

Reviews: