R3765/67G Series Network Analyzer Operation Manual
4.4 Initial Setup
4-3
4.4
Initial Setup
4.4.1
How to Initialize
Operating Procedure
•
Press
[
PRESET
]
key.
The contents of the setup is initialized to the following initial setup value.
4.4.2
Initial Setup Value
Table 4-1 Initial Setup Value (1 of 3)
Function
Initialize method
Power on or preset
*RST
Stimulus
Sweep type
Continuous sweep
Trigger source
Trigger delay
Sweep time
Measurement point
Start frequency
Stop frequency
Frequency display
Fixed frequency of level sweep
Output level
Start level
Stop level
Two-channel interlocking
Program sweep segment
Linear frequency sweep
ON
Internal (FREE RUN)
OFF (0sec)
65.325msec (AUTO)
201
300kHz
3.8GHz (R3765AG/BG/CG)
8.0GHz (R3767AG/BG/CG)
Start/Stop
1GHz
*
1
*
2
*
2
ON
All clear
Linear frequency sweep
OFF
Internal (FREE RUN)
OFF (0sec)
390.325msec (AUTO)
1201
300kHz
3.8GHz (R3765AG/BG/CG)
8.0GHz (R3767AG/BG/CG)
Start/Stop
1GHz
*
1
*
2
*
2
ON
All clear
Response
Dual channel
Active channel
Resolution bandwidth
Selection item of input port
Average
Trace operation
Conversion
Characteristic impedance Z
0
Measurement format
Group delay aperture
Smoothing
Display
Split/Overlap
Label
OFF
CH1
10kHz
*
3
OFF (Number of times 16)
NONE
NONE
50
Ω (75Ω
for OPT12)
*
4
1%
OFF (Aperture 1%)
Data
Overlap
NONE
OFF
CH1
10kHz
*
3
OFF (Number of times 16)
NONE
NONE
50
Ω (75Ω
for OPT12)
*
4
0.01%
OFF (Aperture 0.01%)
Data
Overlap
NONE