R3765/67G Series Network Analyzer Operation Manual
10.1 Preparing for a Performance Test
10-3
Table 10-1 Required Measurement Instrument for Performance Test (3 of 3)
10.1.3 General Note
•
Use an AC power source having a voltage of 90V to 250V and a frequency of 48Hz to 66Hz.
•
When connecting the power supply cable, turn OFF the POWER switch.
•
The R3765/67G series must be tested under the following conditions:
Temperature
: +23
°
C to
±
5
°
C
Relative humidity : 80% RH or less
Free from dust, vibration and noise.
Test items
Measurement Instrument
Remarks
Dynamic
accuracy
• Step attenuator
Variable range: 0 dB to 90 dB
Accuracy
: Within 0.02 dB
HP8496B
(Calibrated under the
national standard)
Refer to
section 10.9
• RF cables
(SMA(m)/SMA(m) 50
Ω
)
×
2
A01253-060
• Conversion connectors
(N(m)/SMA(f))
×
2
HRM-554S
• 3dB fixed attenuators
(SMA(f)/SMA(m))
×
2
AT-103
• 50
Ω
to 75
Ω
conversion adapter
(when OPT 12 or 13 is installed)
(75
Ω
N(m)/50
Ω
N(f))
×
2
11852B
(Agilent Technolo-
gies)
Attenuation
accuracy
(OPT10)
• RF cable
(SMA(m)/SMA(m) 50
Ω
)
A01253-060
• Conversion connectors
(N(m)/SMA(f))
×
2
HRN-554S
• RF cable (For OPT12 or 13)
(N(m)/N(m) 75
Ω
)
8120-6468
(Agilent Technolo-
gies)