A.3.1
Descriptions of Window Components
A-28
Settling:
Enter a settling time.
A wait specified as the settling time is caused at the beginning of a
segment.
RBW:
Select a resolution bandwidth (RBW) from 15 kHz to 10 Hz in the
drop down menu. The selectable resolution bandwidths are as fol-
lows:
15 kHz, 10 kHz, 7 kHz, 5 kHz, 4 kHz, 3 kHz, 2 kHz, 1.5 kHz, 1 kHz,
700Hz, 500 Hz, 400 Hz, 300 Hz, 200 Hz, 150 Hz, 100 Hz, 70 Hz,
50 Hz, 40 Hz, 30 Hz, 20 Hz, 15 Hz, 10 Hz
Aperture:
Sets a smoothing aperture when the group delay display format is
selected.
A.3.1.9
Setting Area for Other Measurement Items
Figure A-33 Measurement Item Setting Area
Each titled field sets the following information:
MemoNo:
Sets a number to be saved to memory (1 to 3).
Save Clr:
Clears the memory saving status.
Save Sv:
Saves to memory.
Disp ON/OFF:
Sets ON/OFF for the display of data saved in memory.
Z0:
Specifies characteristic impedance with a numeric value.
Conv:
Selects a parameter conversion condition from the drop down
menu.
The selectable items are as follows:
OFF:
Does not perform conversion.
Z:TRAN:
Performs impedance conversion using transmission
measurement.
Y:TRAN:
Performs admittance conversion using transmission
measurement.
Z:REFL:
Performs impedance conversion using reflection
measurement.
Y:REFL:
Performs admittance conversion using reflection
measurement.
1/S:
Converts into an inverted S parameter.
SW(ms):
Specifies the sweep time with a numeric value. The unit is always
"msec".