A.3.1
Descriptions of Window Components
A-27
A.3.1.8
Stimulus
Figure A-32 Stimulus
Each titled field sets the following information:
Stimulus:
Specify a segment number from 0 to 1600.
A segment number can be specified when the total sweep point
counts defined for individual segments is 1601 points or less.
When the following items are set, the segment setting becomes
effective.
When deleting a segment, specify 0 in the "Point" field.
However, this operation is effective only for the segment with the
maximum segment number.
Example When segments have been set from 0 to 5, the seg- ment
that can be deleted is only segment 5. If it is nec- essary
to delete segments 1 to 5, delete them one by one in
descending order from 5.
Start:
Specify a start frequency for sweep.
Stop:
Specify a stop frequency for sweep.
Cent:
Specify a center frequency for sweep.
Span:
Specify a frequency span for sweep.
Select a span input unit from the drop down menu, as necessary.
Point:
Specify the number of points for sweep.
Pow1:
Enter a start power level.
For frequency sweep, specify the same value for Pow1 and Pow2.
Pow2:
Enter a stop power level.
For frequency sweep, specify the same value for Pow1 and Pow2.
NOTE:
When the Pow2 check box is selected, power level is output
from Pow1 to Pow2.