A.3.1
Descriptions of Window Components
A-25
DLD button
Displays the DLD function window.
Figure A-30 DLD Function
Meas time:
Displays the DLD measurement time.
Cent [MHz]:
Sets a center frequency where DLD measurement is performed.
Span [ppm]:
Sets a width of two frequencies where DLD measurement is per-
formed.
REF /DIV POS:
Sets the DLD measurement result graph display.
CI[ohm]:
Sets the CI graph display.
∆
F [Hz]:
Sets
∆
F graph.
Start [dBm]:
Sets a DLD measurement start level.
Stop [dBm]:
Sets a DLD measurement stop level.
Points:
Sets the number of DLD measurement points.
RBW [Hz]:
Sets RBW for DLD measurement.
Settling [ms]:
Sets device response time for DLD measurement.
Typ.CI [
Ω
]:
Sets the impedance for the DLD measurement target device.
NOTE:
Electric power is calculated by the impedance set here.
Up & Down:
When selected, the level changes as Start level
→
Stop level
→
Start
level.
By default, it changes as Start level
→
Stop level.