R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
A.2 Error Message
A-3
A.2
Error Message
In this section, the error messages that are displayed while the analyzer is being used are described.
Code
Error message
Remarks
1
Sound demodulation is active.
Turn Sound demodulation off.
2
Vertical scale is set to Linear.
Set the scale to dB/div.
3
Preselector is turned on.
Select manual tuning.
5
Span is set to 0 Hz.
Change the span.
7
Trace mode is set to Blank.
Change to Write mode.
8
Not available for baseband freq.
Move marker before executing.
9
Power measurement is active.
Turn power measurement off.
10
Signal track is active.
Turn Signal Track off.
11
Noise measurement is active.
Turn Noise measurement off.
13
Frequency Counter is active.
Turn Frequency Counter off.
14
Delta marker is not active.
Turn the Delta marker on.
15
External mixer is selected.
Set the mixer to Internal.
17
Not available in Separate-screen mode.
Set to Full-screen mode.
18
View/Blank Trace mode is selected.
Set the Trace to Write mode.
19
Trigger source incorrect.
Set the trigger source to Video/IF.
Set the trigger source to Video or IF.
Change the trigger source to Video or IF.
20
Marker not on selected trace.
Move the marker first.
22
Scale not set to 10dB/div.
Change to 10dB/div first.
23
Parameter is out of range.
25
Calculated power is off the scale.
Calculated result is outside the scale.
Set the reference level to a higher value.
26
Editor is active.
Quit the editor first.