R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
5.2.9 GPIB Command Codes
5-39
Table 5-19 UTIL Key (Utility) (1 of 3)
Function
Listener Code
Talker Request
Panel
Key
Remarks
Code
Output Format
OBW
Measure-
ment
OBW measurement mode
ON
OBW ON
OBWON?
0: OFF
UTIL
OFF
OBW OFF
1: ON
UTIL
Reading the result
–––
OBW?
OBW, Fc
UTIL
A total of 2 outputs
(Both for Frequency)
OBW%
OBWPER *
OBWPER?
Real number (10.0 to 99.8%)
UTIL
Averaging number
AVGOBW *
AVGOBW?
Integer (1 to 999)
UTIL
Use "OFF" when "1" is
set.
Parameter setup
Default
OBWST USR
OBWST?
0: STD (Unused)
UTIL
Manual
OBWST MNL
1: Default
UTIL
Define
→
Default
OBWST DEF
2: Manual
UTIL
Harmon-
ics
measure-
ment
Harmonics measurement
mode
ON
HARM ON
HARMON?
0: OFF
UTIL
OFF
HARM OFF
1: ON
UTIL
Reading the result values
–––
HARM?
Number of sets n<DLM>
Frequency1, Level1<DLM>
...
Frequency n, Level n<DLM>
UTIL
n = HRMNUM?
<DLM>=Delimiter
Harmonics Number
HRMNUM *
HRMNUM?
Integer
UTIL
Fund Frequency
HRMFND *
HRMFND?
Frequency
UTIL
Fund Frequency ON
HRMFND ON
HRMFNDON? 0: OFF
UTIL
OFF
HRMFND OFF
1: ON
UTIL
Spurious Spurious measurement ON
SPURI ON
SPURION?
0: OFF
UTIL
OFF SPURI OFF
1: ON
UTIL
Reading the result values
–––
SPURI?
Number of times the measurement
table n<DLM>
m1<DLM>
f1, l1, j1 <DLM>
...
fm1, lm1, jm1<DLM>
m2<DLM>
f1,l1,j1<DLM>
...
fm2, lm2, jm2<DLM>
. . .
. . .
mn<DLM>
f1, l1, j1<DLM>
...
fmn, lmn, jmn<DLM>
UTIL
Number of times the
measurement table
must be repeated (0 to
10)
m: Number of times
the spurious must be
repeated (0 to 10)
f: Spurious frequency
l: Spurious level
j: Test conclusions
<DLM>=Delimiter
Table selection
SPRTBL *
SPRTBL?
Integer (1 to 3)
UTIL
Saving the table information SPRSV
–––
–––
UTIL
Loading the table
information
SPRLD
–––
–––
UTIL
Entering data in the table
SPRIN *
–––
* Input number (integer), start
frequency, stop frequency,
RBW and limit value
UTIL
Deleting the table
SPRDEL
–––
–––
UTIL
Deleting all data in the