5.2.9 GPIB Command Codes
5-24
Table 5-3 CAL Key (Calibration)
Function
Listener Code
Talker Request
Panel Key
Remarks
Code
Output Format
Calibration
Cal ALL
CLALL
–––
–––
CAL
Total Gain Cal.
CLGAIN
–––
–––
CAL
Input ATT Cal.
CLATT
–––
–––
CAL
IF Step AMP Cal.
CLSTEP
–––
–––
CAL
RBW Switching Cal.
CLRBW
–––
–––
CAL
Log Linearity Cal.
CLLOG
–––
–––
CAL
Amplitude MAG Cal.
CLMAG
–––
–––
CAL
PBW Cal.
CLPBW
–––
–––
CAL
Calibration level
CLN *
CLN?
Level
CAL
Cal 10 M Reference Coarse
CLCREF *
CLCREF?
Integer (0 to 255)
CAL
Cal 10 M Reference Fine
CLFREF *
CLFREF?
Integer (0 to 255)
CAL
Cal 10 M Reference Default
CLDREF
–––
–––
CAL
Saves the cor-
rection value.
Cal 10 M Reference Store
CLSREF
–––
–––
CAL
Initializes the
correction
value.
Freq Corr ON
FC ON
FC?
0: OFF
CAL
OFF
FC OFF
1: ON
CAL
Cal Corr ON
CC ON
CC?
0: OFF
CAL
OFF
CC OFF
1: ON
CAL