![ADVANTEST R3267 series Operation Manual Download Page 220](http://html1.mh-extra.com/html/advantest/r3267-series/r3267-series_operation-manual_2858766220.webp)
R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
3.3.5 CONFIG Key (Configuration)
3-35
Day/Month/Year:
Displays in the order of a day of the week, day, month
and year.
Trace Point 1001/501
Toggles the trace point on the horizontal axis between 1001 and
501.
1001:
Sets the trace points to 1001.
501:
Sets the trace points to 501.
Display ON/OFF
Toggles the annotation display function on or off.
ON:
Displays the annotation.
OFF:
Removes the annotation.
more 1/2
Displays the Config (2) menu.
Selftest
Displays the Selftest menu.
Execute Selftest
Executes the selftest.
Power ON Count
Displays the number of times the spectrum analyzer is turned on,
accumulated total of powerup time.
ATT switch Count
Displays the total switching counts for each internal cells of the
attenuator.
Test Mode Exit
Terminates the self-test mode. All settings are reset to their intial
values and the spectrum analyzer stops sweeping.
Revision
Displays the software versions and the options implemented in
the analyzer.
more 2/2
Returns the Config (1) menu.