R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
3.3.4 CAL Key (Calibration)
3-29
Fine
Allows you to enter a fine correction data to the 10 MHz reference
frequency.
Store
Saves the correction data corresponding to the 10 MHz reference
frequency that have previously been modified.
Default
Resets the coarse and fine correction data that are previously en-
tered to the factory defaults.
Freq Corr ON/OFF
Toggles the frequency correction function on or off.
ON:
Frequencies are corrected based on characteristics set at
the factory.
OFF:
Turns the frequency correction function off.
more 2/2
Displays the CAL(1) menu.