R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
3.3.3 B Key (Trace B)
3-26
Min Hold B
Allows you to set the number of times the sweep Min Hold is per-
formed. Once sweeping has begun, the minimum result for each
sweep is kept and displayed until the set count is reached.
Normalize B
Displays the Normalize B menu.
Normalize B ON/OFF
Toggles the Normalize function on or off.
ON:
Corrects for the level using the normalization data.
OFF:
Turns the Normalize function off.
Normalize with Store Corr.
This command obtains normalization data and turns the normal-
ization function ON. The waveform data that is displayed on the
screen is used for obtaining the normalization data.
Display Line Pos.
Displays the display line and allows you to set the position of the
display line.
A
←→
←→
←→
←→
B
Switches the data saved in memory A with the data saved in mem-
ory B, and memory B data with memory A.
Power Average B
Displays the trace averaged in units of watt, using data in dBm.
Where Pavg is the result of averaging the power; Pin is
Nth measurement data for one point (1 to 1001); and n
is the number of averaging (or number of sweeps)
Trace B Detector
Displays the Detector B menu.
Normal
Sets the normal detector mode which automatically detects posi-
tive or negative peaks for each trace point.
Positive
Sets the positive peak detector mode.
Negative
Sets the negative peak detector mode.
Sample
Sets the sample detector mode.
P
AVG
= 10log [
× ∑
10
1001
Pin
i = 1
(
)
n
1
10