45
Appendix C
C.3 D/A Calibration (PCI-1710U/HGU)
In a way similar to the A/D Calibration, the D/A calibration tool in the
Advantech Device Manager leads you through the whole D/A calibration
procedure.
1.
Measure the DA0 output with a 4 1/2-digit digital meter. Adjust the
VR4 until the output value equals 4.9988V.
Note:
Using a precision voltmeter to calibrate the D/A
outputs is recommended.
Summary of Contents for PCI-1710HGU
Page 1: ...PCI 1710U Series 12 bit Multifunction Cards with Universal PCI Bus User Manual...
Page 12: ...PCI 1710U Series User Manual 6 Figure 1 1 Installation Flow Chart...
Page 24: ...PCI 1710U Series User Manual 18...
Page 35: ...2 APPENDIX A Specifications...
Page 41: ...35 Appendix A...
Page 42: ...PCI 1710U Series User Manual 36...
Page 43: ...2 APPENDIX B Block Diagrams...
Page 44: ...PCI 1710U Series User Manual 38 Appendix B Block Diagrams...