![Advanced Instruments OsmoTECH XT User Manual Download Page 16](http://html.mh-extra.com/html/advanced-instruments/osmotech-xt/osmotech-xt_user-manual_2856049016.webp)
17
OsmoTECH XT Single-Sample Micro-Osmometer User Guide
Chapter 1
Chapter 1:
Installation and setup
To set up your instrument properly, it is important to read and carefully follow the steps in this section. Also read
before attempting to run your instrument.
1.1 Selecting a location for the instrument
Select a location for the OsmoTECH XT based on the
following space and electrical requirements.
OsmoTECH XT dimensions
•
Not including test kit:
15" D x 14" W x 10" H
(38 cm D x 36 cm W x 25 cm H)
•
Including test kit:
15" D x 14" W x 11.5" H
(38 cm D x 36 cm W x 29 cm H)
Recommended clearance
Allow 6 inches (15 cm) of clear space on each side of the
instrument.
Do not block the exhaust fan airflow on the back panel
or underneath the instrument.
Electrical requirements
Your instrument must be located within five
feet of a properly grounded, three-prong
electrical outlet capable of continuously
supplying 0.5 amperes at 100–240 VAC
(50–60 Hz).
If the instrument is not grounded properly, its
operation may be impaired and a safety hazard
may exist. Verify that the outlet is appropriate
before operating your instrument.
Do not disable the grounding plug.
1.2 Unpacking the shipment
The following table lists the items contained in your
shipment.
Qty
Description
1
OsmoTECH XT Single-Sample Micro-Osmometer
1
20-
µ
L
Ease-Eject™
Sampler
1
Power cord
1
User Guide on USB drive
NOTE:
Advanced Instruments cannot guarantee the
stated instrument performance specifications
and accuracy of test results unless Advanced
Instruments brand consumables, standards, and
controls are used with the instrument.
•
For information on obtaining Advanced Instruments
brand items, see
Supplies, standards and controls
•
For information about Advanced Instruments references
and controls, see
.
NOTE:
Use of consumables, standards, or controls from
manufacturers other than Advanced Instruments
is not recommended and may adversely affect
system calibration, performance, operation, and
accuracy of test results.
Carefully unpack the instrument, accessories, and
supplies and inspect them for shipping damage. Use the
enclosed packing list to verify that all items are received.
1.
If any item on the packing list appears to be missing
from your shipment, please search carefully through
and under all packing materials. If you don't find the
item, notify your receiving department immediately.
Advanced Instruments can only be responsible for
items reported missing within 10 days of a shipment
arrival.
2.
Save the shipping boxes and packaging material for
future use if transport of the instrument becomes
necessary.