Admag SE100DJ/EJ User Manual Download Page 8

IM 1E10D0-01E

3-3

3.  INSTALLATION

(12) Grounding

IMPORTANT

Improper grounding can have an adverse affect
on the flow measurement. Please ensure that
the instrument is properly grounded.

The electromotive force of the magnetic flow tube is
minute and it is easy to be affected by noise. And also
that reference electric potential is the same as the
measuring fluid potential. Therefore, the reference
electric potential (terminal potential) of the Flow Tube
and the Converter/Amplifier also need to be the same
as the measuring fluid. And moreover, that the poten-
tial must be the same with ground.

Please be sure to ground according to Figure 3.1.7.

600 V vinyl insulated
electric cable
(2mm

2

 or larger)

Earth ring

F030107.EPS

Grounding 

resistance 100

 or less

Note: See “3.4.1 Grounding” for 

information on grounding.

In case earth rings are not used.
(Available only for metal piping)

In case earth rings are used.

Figure 3.1.7

Grounding

3.2 Handling Precau-

tions

WARNING

The Magnetic Flow tube is a heavy instrument.
Please be careful to prevent persons from
injuring when it is handled.

3.2.1

General Precautions

(1) Precaution for Carrying

The Magnetic Flow Tube is packed tightly. When it is
unpacked, please give attention to prevent damages to
the flow tube. And to prevent the accident during carry
to the installing location, please carry it near the
location keeping packed as it delivered.

CAUTION

In case the Magnetic Flow Tube without eye-bolt
lifts up, please refer to Figure 3.2.1. Please
never lift up by using a bar through the flow
tube. It damages liner severely.

F030201.EPS

Figure 3.2.1

Vertical Lifting Sling Rigging Method

(2) Precaution for Shock

CAUTION

Care should be taken not to drop the flow tube
or subject it to excessive shock. This may lead
to liner damage which will cause inaccurate
readings.

(3) Flange Protection Covers

IMPORTANT

Please keep the protection cover (ex. corrugated
paper or anything possible to protect) attached
with flange except when mounting to the pipe.

(4) Terminal Box Cover

IMPORTANT

Please never leave the terminal box cover open
until wiring to prevent insulation deterioration.

NOTE

The terminal box cover is locked by special
screw.  In case of opening the terminal box
cover, please use the Hexagonal Wrench
attached.

Summary of Contents for SE100DJ/EJ

Page 1: ...User s Manual Y okogawa Electric Corporation Models SE100DJ EJ SE200DJ EJ and SE300DJ EJ Magnetic Flow Tube IM 1E10D0 01E IM 1E10D0 01E 8th Edition...

Page 2: ...recautions 3 3 3 2 2 Flow Tube Piping 3 4 3 3 Mounting 3 4 3 3 1 Nominal Diameter 15mm 0 5 to 40mm 1 5 Wafer Type 3 4 3 3 2 Nominal Diameter 50 mm 2 to 200 mm 8 Wafer Type 3 6 3 3 3 Nominal Diameter 1...

Page 3: ...recautions The following general safety precautions must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions or with specific WARN...

Page 4: ...self because the surface of body and case reach a high temperature When removing the instrument from hazardous processes avoid contact with the fluid and the interior of the flow tube All installation...

Page 5: ...e Precautions In case the instrument is expected to be stored over a long term please give attention to the followings The instrument should be stored in its original packing condition The storage loc...

Page 6: ...ucer pipe Expander pipe Tee 90 degrees bent Various types of valves 5 D or more 2 D or more 2 D or more 2 D or more 10 D or more 10 D or more 5 D or more 5 D or more Do not care to 0 D Do not care to...

Page 7: ...uid is always filled in the pipes The Vertical Mounting is effective for fluids that is easily separate or slurry settles within pipes In this case please flow a fluid from bottom to up GOOD GOOD BAD...

Page 8: ...prevent persons from injuring when it is handled 3 2 1 General Precautions 1 Precaution for Carrying The Magnetic Flow Tube is packed tightly When it is unpacked please give attention to prevent dama...

Page 9: ...elding scrap or wood chips may exist Please remove them by flushing piping before mounting the flow tube Slant Mis alignment F030202 EPS Figure 3 2 2 Slant and Mis alignment of Flow Tube Piping 3 3 Mo...

Page 10: ...pieces Please use appropriate bolts and nuts according to Process connection Horizontal Mounting Vertical Mounting F030301 2 EPS Figure 3 3 1 Mounting Procedure Size 15 mm 0 5 to 40 mm 1 5 Table 3 3 1...

Page 11: ...es two for each and lower adjacent holes of both flanges Refer to Figure 3 3 2 Please give attention to the nominal size and flange ratings of the centering devices Refer to Table 3 3 5 3 Positioning...

Page 12: ...has equal elasticity T030303 EPS Table 3 3 4 Tightening Torque Values for PVC Piping in N m kgf cm in lbf Size mm inch JIS 10K JIS 20K ANSI 150 ANSI 300 DIN PN10 50 2 80 3 100 4 150 6 200 8 8 2 84 73...

Page 13: ...termi nal box In case the measuring fluid flows against the arrow direction please refer to the section Reversing Flow Direction in the Instruc tion Manual of SE14 Magnetic Flow Converter Flange Flow...

Page 14: ...363 531 65 0 93 0 663 949 575 823 125 1 151 5 1280 1550 1112 1347 153 5 180 9 1570 1850 1365 1608 94 0 125 0 959 1275 832 1106 135 9 164 3 1390 1680 1208 1460 154 5 199 5 1580 2040 1373 1773 151 5 192...

Page 15: ...er if it is too short Please replace the short cable with the cable which is appropriate length wholly The all cable ends are to be provided with round crimp on terminal The signal cables must be rout...

Page 16: ...cables JIS C3401 or Polyvinyl chloride insulated and sheathed portable power cables JIS C3312 or equivalents Outer Diameter 6 5 to 12mm in diameter 10 5 to 11 5 mm for waterproof gland ECG ECU Nominal...

Page 17: ...terminal C and terminal A B Between terminal A and terminal B Between terminal EX1 and ground Between terminal EX2 and ground All insulation measurements must be performed with a 500V megger Insulati...

Page 18: ...converter Please refer to Instruction Manual of SE14 magnetic flow converter IMPORTANT Be sure to disconnect the cables at the terminals of the flow tube when checking 1 Excitation Coil Check Check th...

Page 19: ...and earth ring for optional code FRG Non asbestos joint sheet sheathed with fluoro resin PTFE between earth ring and process flange for optional code BSF Other gaskets between flow tube and earth rin...

Page 20: ...00 150 200 250 300 350 400 0 1909 0 5302 1 3572 2 1206 5 429 8 483 19 086 33 93 53 02 76 35 103 91 135 72 Size 6 361 17 671 45 23 70 68 180 95 282 74 636 1 1 130 9 1 767 1 2 544 6 3 463 4 523 SI Units...

Page 21: ...for 80 to 150mm because of same mating dimensions Note 2 Only ANSI1 2NPT electrical connection is available for FM or CSA explosion proof type JIS G1 2 electrical connection is not available for any...

Page 22: ...ing Treatment Degreased cleansing treatment Packing with desiccant Coating is changed to three layer coating Urethane coating on two layer Epoxy coating X2 K5 GOS Calibration Certificate for GOST Only...

Page 23: ...8 1 H H1 L 1 111 4 4 48 1 9 80 5 3 2 Lining Outside diameter Height Fluorocarbon PFA 86 3 4 D D D Ground Terminal SD2d eps 1 When no earth ring is selected the face to face length is shorter by approx...

Page 24: ...11 5 11 5 11 3 13 0 10 5 6 4 6 4 6 1 6 2 5 4 5 5 4 9 4 9 0 7 0 7 1 1 2 0 2 4 3 1 Flange outside dia Dia of holes Number of holes Thickness Except P T Earth ring D C 10 1 Lining Fluorocarbon PFA Fluor...

Page 25: ...ole pitch 22 5 22 5 22 5 22 5 22 5 22 5 22 5 22 5 22 5 45 45 45 150mm 6in 200mm 8in Flange Type Unit mm Approx inch Ground Terminal 80 5 3 2 48 1 9 86 3 4 111 4 4 20 5 0 8 68 2 7 H1 H2 L 1 T 2 H C D N...

Page 26: ...10 16 400 16 22 0 9 26 1 431 8 17 450 8 18 25 4 1 31 7 1 2 414 16 3 23 0 9 40 1 6 59 2 3 34 1 3 36 1 4 34 1 3 34 1 3 400 15 7 25 1 44 1 7 430 16 9 27 1 1 16 16 10 12 12 16 12 96 104 87 88 87 87 91 SE3...

Page 27: ...nstallation WARNING All wiring shall comply with local installation requirements and local electrical code In hazadous locations the cable entry devices shall be of a certified flameproof type suitabl...

Page 28: ...ure 70 C 85 C 120 C 150 C Minimum Process Temperature 40 C 40 C 40 C 40 C Enclosure NEMA 4X Ambient Temperature 20 to 60 C Maximum Working Pressure 4MPa SE115E to SE205E 2MPa SE208E to SE220E Flange r...

Page 29: ...of Certification 6 4 SAA 1 Technical Data SA Certificate No AUS Ex 3764X Type of Protection Ex d m ia II C T6 T3 Enclosure Type IP67 Excitation Circuit 41Vmax 6 6 25Hz Electrode Circuit Um 250V ac dc...

Page 30: ...02D E are not attached CE mark of PED because they do not come under CE marking of PED 2 Installation WARNING Please tighten the bolts for piping joint according to the prescribed torque values Please...

Reviews: