Calibration & Utilities
•
75
6
Calibration & Utilities
In data acquisition process, how to calibrate your measurement devices
to maintain its accuracy is very important. Users can calibrate the
analog input and analog output channels under the users' operating
environment for optimizing the accuracy.
The software CD provides two utility programs. They are 9114util.exe
which provides three functions, System Configuration, Calibration, and
Functional Testing, and I_eeprom which is used to enable or disable
interrupt of PCI-9114 board. The utility programs are described in the
following sections.
6.1 Calibration
6.1.1
What do you need
Before calibrating your PCI-9114 card, you should prepare some
equipments for the calibration:
•
Calibration program: Once the program is executed, it will guide
you to do the calibration. This program is included in the
delivered package.
•
A 51/2 digit multi-meter (a 61/2 meter is recommended)
•
A voltage calibrator or a very stable and noise free DC voltage
generator.
Summary of Contents for NuDAQ PCI-9114DG
Page 1: ...NuDAQ P C I 9 1 1 4 D G H G Enhanced Multi Functions Data Acquisition Card User s Guide ...
Page 2: ......
Page 5: ......
Page 6: ......
Page 10: ...iv Contents Product Warranty Service 83 ...
Page 12: ......
Page 21: ...Installation 9 2 3 PCB Layout of PCI 9114 ...
Page 31: ...Registers Format 19 CN4 is MSB and CN0 is LSB ...
Page 36: ......
Page 50: ......
Page 56: ...44 C C Library the digital value Return Code ERR_NoError ...
Page 79: ...C C Library 67 ...
Page 94: ......