ADInstruments Transducer Series Quick Start Manual Download Page 1

AD

INSTRUMENTS

MLT1115 Galvanic Oxygen Electrode

Transducer Series

Description

The MLT1115 Galvanic Oxygen Electrode is 

designed for measuring the concentration of oxygen 

in aqueous solution.

Operation

Connect the electrode to the input connector of any PowerLab (or other recording 

device having an input impedance of at least 1 MΩ). The unit may be connected directly 

into a BNC input or alternately to a Pod Port via the supplied MLAC22 BNC to DIN 

Smart Adaptor. The output in air-saturated deionized water should be 20 – 35 mV at 

25 °C. The output will be larger at higher temperatures or in water saturated with pure 

oxygen or Carbogen (5% CO

2

, 95% O

2

).

Note: Constant temperature must be maintained to obtain a meaningful signal.

Principle of Operation

The electrode comprizes a platinum cathode and a lead 

anode connected via an internal 7.5 kΩ resistor. As oxygen 

diffuses across the membrane and into the electrolyte 

solution inside the electrode, the following reactions occur:
 O

2

 + 2H

2

O + 4e

-

 →  4OH

-

 

at cathode

  2Pb + 4OH

-

 

→ 2Pb(OH)

2

 + 4e

-

  at anode

 O

+ 2H

2

O + 2Pb  →  2Pb(OH)

2

 overall

For every molecule of oxygen consumed at the cathode, 

four electrons are transferred from the anode. The potential 

across the internal 7.5 kΩ resistor is proportional to this 

current flow therefore, to oxygen concentration.
The rate of oxygen consumption by the electrode is given 

by the formula:
 d[O

2

]/dt = E/(4000RF)

where:

 d[O

2

]/dt is the rate of oxygen consumption in mol/s

  E is the output signal (potential) in mV
  R is the value of the internal resistor, 7.5 kΩ
  F is the Faraday constant, 96486.7 A·s/mol

Membrane

Cathode

Cover cap, 

tip

Anode

Cover cap, 

cylinder

Summary of Contents for Transducer Series

Page 1: ...meaningful signal Principle of Operation The electrode comprizes a platinum cathode and a lead anode connected via an internal 7 5 kΩ resistor As oxygen diffuses across the membrane and into the electrolyte solution inside the electrode the following reactions occur O2 2H2 O 4e 4OH at cathode 2Pb 4OH 2Pb OH 2 4e at anode O2 2H2 O 2Pb 2Pb OH 2 overall For every molecule of oxygen consumed at the ca...

Page 2: ...de is immersed in stirred deoxygenated water a steady level must be obtained for a couple of minutes Transfer the electrode to a stirred air saturated solution while continuing to record the signal until a steady level is obtained in LabChart Stop recording Use the Units Conversion feature of the LabChart software see the LabChart User s Guide to assign values of zero and 100 O2 saturation to the ...

Page 3: ...n The filling solution should be replaced periodically Troubleshooting Q The probe when in air saturated water displays a zero signal A Check that all cable connections are correct Check that the probe is filled with electrolyte filling solution Q The maximum signal in air saturated water is less than 20 mV A The output signal will be smaller at temperatures less than 25 C If the membrane is dirty...

Page 4: ...tatement of Intended Use on our website or in Getting Started with PowerLab before use The electrode filling solution is caustic and contains sodium hydroxide Wear rubber gloves and eye protection when refilling the electrode If skin or eye contact occurs wash the affected area immediately with running water Specifications Probe type Galvanic Cathode Platinum Anode Lead Membrane PTFE Teflon Output...

Reviews: