7
Maintenance
When maintaining (cleaning, sterilizing, etc.) the stimulator, be
sure to turn the power off and unplug the power cable from the
power outlet first. Failure to do so may result in an electric
shock or malfunction.
CAUTION
After Use Inspection
Be sure to conduct the following after-use inspections after finishing stimulator
use. These inspections should be performed to ensure smooth and normal
operation next time.
After-Use Inspection tasks
Item
Description
Operating
condition
Check if the device is in normal working order.
Check for dirt, scratches and damages on the exterior.
State and
storage
Check if the electrodes are clean.
Wipe off any moisture and dry the device thoroughly.
Check to see if the device was turned off.
Check if any chemicals, water or other remain on the device.
Check if the device is properly stored.
Cleaning
Device Exterior
Note: Turn off the power and unplug the device before maintenance
(cleaning, disinfection, etc.) to prevent malfunction and electric
shock.
Use a soft cloth moistened in a solution of neutral detergent to wipe away any
dirt. Then remove moisture with a dry soft cloth.
Note: Never use thinner, acetone, benzene or other organic solvents as
they may damage the paint or resin parts.
After using water (cleanser) for cleaning, remove any remaining
moisture with a thoroughly wrung out cloth making sure that no
liquid enters the interior of the device.
Electrodes
Cationic soap and water will be enough for normal cleaning.
Dry all metal parts thoroughly before storing to prevent rust.
Do not use alcohol to clean electrode leads (or extension cables) as it may cause
the vinyl coating to harden.