5
Output setup for constant-voltage stimulation
1.
Set the electronic stimulator in operating mode and set
the output voltage to maximum.
2.
Use the +/– switch (Stimulator output display polarity
switch) to select the output polarity to be displayed on
the OUTPUT MONITOR.
+ : Displays the positive pulse of stimulator output
– : Displays the negative pulse of stimulator output
* This setting is not related to the setting in step 5.
3.
Set the MODE knob to the required VOLTAGE range.
4.
While observing the OUTPUT MONITOR, turn the
FINE adjuster to adjust the output voltage.
5.
Use the
switch (output polarity switch) to select
the output polarity from this device.
: Red terminal b
: Red terminal becomes –
6.
Set the STIM/OFF switch to “STIM” to start stimulation
output.
7.
To multiply the set voltage, turn the VOLTAGE MAG
knob to the target voltage multiplier.
Output setup for constant-current stimulation
1.
Set the electronic stimulator in operating mode and set
the output voltage to maximum.
2.
Set the MODE knob to the required CURRENT range.
3.
While observing the OUTPUT MONITOR, turn the
FINE adjuster to adjust the output current.
4.
Use the
switch (output polarity switch) to select
the output polarity.
: Red terminal b
: Red terminal becomes –
5.
Set the STIM/OFF switch to “STIM” to start stimulation
output.
switch
S T I M / O F F
switch
VOLTAGE MAG
knob
MODE knob
+/– switch
OUTPUT MONITOR
FINE adjuster
switch
STIM/OFF switch
MODE knob
OUTPUT MONITOR
FINE
adjuster