9 Fault
A fault of the ADICOS GSME is indicated by illumination of the yellow signal LED »Fault« as well
as by deactivation of the »Fault« signal relay.
The following fault states and/or causes are recognized by the detector:
Fault
Description
Measure
Undervoltage/
power outage
Operating voltage drops below
20 V (including temporarily)
Check supply voltage;
Check wiring;
Use stronger or additional
power supply;
Self-resetting
Checksum
ROM / EEPROM
The checksum is incorrect after
a new start or reset.
Detector replacement
Heating temperature
One or several heating tempera
-
tures do not reach the target val
-
ue.
Use detector insulation;
Self-resetting
Test fault
A test fault was triggered using
the ADICOS service software.
Cancel test fault using
ADICOS service software
Sensor fault
The conductance of at least one
sensor is exceeded resp. under
-
cut.
Detector replacement
Start-up sequence
The detector always remains in
fault, until the end of the start-
up process.
Self-resetting
Use the ADICOS service software for a detailed analysis of the fault states!
The ADICOS GSME does not offer automatic contamination detection for the
sintered metal filter.