94
5.2.4 Pressure switch
1. Connection
Connect the calibrator and the pressure switch as shown in Figure 3-20;
2. New task
Select the “pressure switch” icon
in the test type list, and input the information of the pressure switch.
Table 5- 6 Task mode pressure switch information
Item
Effective value
Description
Name
Letter, numeral, symbol
Name of the pressure sensor
Serial number
Letter, numeral, symbol
Serial number of the pressure sensor
Model
Letter, numeral, symbol
Model of the pressure switch
Pressure type
Gauge pressure, absolute
pressure, differential pressure
Pressure type of the pressure switch, check the support conditions
according to the calibrator model
Input
Depends on the pressure switch Pressure range of the pressure switch
Accuracy
0.5%, 1%, 1.5%, 2%, 2.5%, 4%,
and custom
Precision of the pressure switch. The numeral input in the custom
option is the precision level of the pressure switch
Set point
Depends on the range of the
pressure switch
Operating point of the pressure switch
Action type
Normally closed, normally open
Action type of the pressure switch
Switch type
Mechanical switch, NPN switch,
Type of the pressure switch
Summary of Contents for 761A
Page 1: ...761A Automated Pressure Calibrator...
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Page 19: ......
Page 20: ......
Page 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Page 33: ...13 LAN Interface USB port Power supply input Label Battery...
Page 47: ...27 Figure 3 3 Gauge pressure outlet...
Page 151: ...131...