Table 10 Pressure measurement settings
Subject
Valid Value
Description
Resolution
4/5/6
Resolution of current measurement channel
Pressure type
GP/ AP
Current measurement channel pressure type
Measurement frequency
1~10Hz
Data acquisition frequency of the current measurement
channel
Stability enable
Enable/disable
Whether to turn on the stability judgment function
Stability value
Pressure value or percentage of
range
A value for judging whether the pressure is stable. If the
measured value fluctuates less than this value within the
stable time, the calibrator will judge as stable
Stabilization time
Number
Time to judge the pressure stability
Tare enable
Enable/disable
Enable or disable tare function
Tare value
Pressure value
Correction value of the tare function
3.3.1 Filter
Refer to the section 3.1.6.
3.3.2 Module information
The basic module information, including the serial number, range, pressure type and firmware information.