1.2.2 Signal measure specification (environmental temp.: 20±10
℃
, 1 year accuracy)
Table 2 Signal measure specification
Specification
Range
Accuracy
Resolution
Note
RTD Measurement
Accuracy
0~400Ω
0.01%RDG+0.005%FS[1]
1mΩ
Excitation current: 1 mA
Voltage Measurement
±300mV
0.015%RDG + 0.005%FS
1uV
Impedance:
>100MΩ
±30V
0.015%RDG + 0.005%FS
0.1mV
Impedance:
>1MΩ
Current Measurement
±30mA
0.015%RDG + 0.005%FS
0.1uA
Impedance: <
40Ω
Frequency Measurement
0.01~50000Hz (auto range)
0.005% RDG + 2 last digit
6-digit auto-
resolution
Min threshold voltage: 2.5V
Units: Hz, kHz, MHz, CPM, CPH, s, ms, us
Switch On-Off
Measurement
Inspection voltage: (3 ~ 30)V
Response speed: < 10ms, supports wet and dry switch
Pulse Count
0 ~ 9999999, optional rising edge and falling edge Min threshold voltage: 2.5V
Loop Power
20V ± 10%, max output impedance: 320Ω, max load current:
25mA
Note 1
:
When the environmental temperature is (-20~-10)
℃
, the temperature coefficient is:
(1)Measure of voltage, current and RTD
:
±5ppmFS/
℃
.
Note 2
:
Input characteristics:
(1) Voltage range
:
-300~300mV input impedance >100 M
Ω;