8250A Optical Power Meter Operation Manual
3.4.2 How to Initialize
3-8
3.4.2
How to Initialize
For more information on the parameters to be initialized, refer to Table 3-1.
1. How to initialize all parameters to the factory set values.
•
Press the
[POWER]
switch while pressing the
[UP]
and
[PREV]
keys to turn on the power.
•
“F INI” is displayed and all parameters are initialized to the factory set values.
The USER-0 to USER-3 areas of the built-in flash memory are also initialized.
2. How to initialize all parameters except for common parameters. (By key operations)
•
Select the menu item "P-LD" (parameter load) by pressing
[MENU]
and then
[PREV]
or
[NEXT]
.
•
Select "dFLt" by pressing
[UP]
or
[DOWN]
.
•
Press
[ENTER]
. “DONE” is displayed and all parameters except for common parameters are
initialized.
The USER-0 to USER-3 areas of the built-in flash memory are not initialized.
•
If initializing the USER-0 to USER-3 areas of the built-in flash memory, select "dFLt" in the
menu item "P-SV" (parameter save) and execute.
3. How to initialize all parameters except for common parameters. (By remote operations)
•
Send either the *RST command or RL command from GPIB or USB controller.
The USER-0 to USER-3 areas of the built-in flash memory are not initialized.
•
Send the CL command if initializing the USER-0 to USER-3 areas of the built-in flash mem-
ory.