6247G/6247C DC Voltage Current Source/Monitor Operation Manual
9. SPECIFICATIONS
9-1
9.
SPECIFICATIONS
All accuracy specifications are guaranteed for one year at a temperature of 23
5
C and a relative humidity
not exceeding 85%.
9.1
Source and Measurement
However, the measurement resolution with integration time of 100
s, 500
s or S/H (Sample Hold) is as
follows:
Range
Source range
Setting
resolution
Measurement range
Measurement
resolution
Voltage source/
measurement
range
5 V
-1.0000 to
5.0000 V 100
V
-1.00999 to +5.00999 V 10
V
50 V
-10.000 to
50.000 V 1 mV
-10.0999 to +50.0999 V 100
V
250 V
-15.00 to
250.00 V
10 mV
-15.099 to + 250.999 V
1 mV
Current source/
measurement
range
3
A
0 to
3.2000
A
100 pA
0 to
3.20999
A
10 pA
30
A
0 to
32.000
A
1 nA
0 to
32.0999
A
100 pA
300
A
0 to
320.00
A
10 nA
0 to
320.999
A
1 nA
3 mA
0 to
3.2000 mA
100 nA
0 to
3.20999 mA
10 nA
30 mA
0 to
32.000 mA
1
A
0 to
32.0999 mA
100 nA
300 mA
0 to
320.00 mA
10
A
0 to
320.999 mA
1
A
Resistance
measurement
range
Determined by
voltage range/
current range
calculations
-
-
0
to 125 G
Minimum
30
Integration time
100
s
500
s
S/H (100
s)
Measurement resolution (digits)
10
2
10