6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.8 Source and Measurement Timing
5-47
5.2.8.3
Integration Time and Measurement Time
The measurement time (Tm) is calculated from the integration time (Tit) and the internal processing time
(Tk) according to the following formula:
Tm = Tit + Tk
The integration time (Tit) can be selected between 100
s to 200 ms.
The internal processing time (Tk) is determined by the source mode and memory store mode settings as
below.
When the memory store mode is OFF or Normal-ON, the following processing times are added by NULL
calculation, scaling calculation, Max/Min calculation, and comparator calculation.
NULL calculation ON:
Approx. 0.2 ms
Scaling calculation ON:
Approx. 1 ms
Max/Min calculation ON:
Approx. 1 ms
Comparator calculation ON:
Approx. 15 ms
In measuring resistance:
Approx. 1 ms
(Example) DC source mode, integration time: 1 PLC (50 Hz), memory store: Normal-ON,
NULL calculation: ON, scaling calculation: ON, Max/Min calculation: ON,
comparator calculation: ON
In these conditions, the measurement time is calculated as follows:
Tit = 20 ms
Tk = 4 + 0.2 + 1 + 1 + 15 = 21.2 ms
Tm = Tit + Tk = 41.2 ms
Source mode
Memory store
Tk [ms]
DC
Pulse
OFF
Approx. 4
Normal-ON
Approx. 4
Burst-ON
Approx. 1
DC sweep
Pulse sweep
OFF
Approx. 5
Normal-ON
Approx. 5
Burst-ON
Approx. 1