6247G/6247C DC Voltage Current Source/Monitor Operation Manual
4.2.10 OPR/SUSPEND Key (Operating/Suspend)
4-17
4.2.10
OPR/SUSPEND Key (Operating/Suspend)
OPR key:
Switches between Operate and Suspend.
Operate:
Turns the output status ON and the OPR indicator turns ON.
Displays measurement values when the measurement is ON. (Not
displayed when the measurement is OFF.) The operation varies
depending on the source mode.
DC:
Generates a setting value and displays a measurement
value.
PLS:
Generate a pulse and displays a measurement value.
DC-SWP and PLS-SWP:
Generate a bias value and waits for measurement.
Produces sweep data and waits for trigger input. The
trigger input starts the sweep and displays a
measurement value.
Suspend:
Outputs the suspend voltage without turning OFF the output
relays. The OPR indicator blinks. The suspend voltage is dis-
played and the header indicates the suspend status.
HZ:
High impedance status
LZ:
Low impedance status
SUSPEND key (SHIFT):
Sets Suspend in Standby or Operate status.
NOTE:
Sweep data is produced when the output status is switched as follows in the DC sweep or pulse sweep
mode:
• From Standby to Operate
• From Standby to Suspend
• Switched to Operate after changing the sweep parameters during Suspend.
"Setting" is displayed during sweep data creation. After the sweep data creation finishes, the internal
processing for Operate starts.
4.2.11
SHIFT/LOCAL Key (Shift Mode/Local)
SHIFT
key (in normal operation):
Functions as SHIFT key, and the SHIFT indicator goes ON.
In the shift status, function names printed in blue characters on the
panel are enabled. Press the SHIFT key again to release the shift
status.
LOCAL
key (in remote control):
Releases remote control.
The RMT indicator goes OFF.
NOTE:
Remote control is not released if the LLO (Local Lock Out) command is set through the GPIB or USB
interface.