6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-38
Cali
bra
tion
Calibration switch CAL0
OFF (Exits the calibration mode.)
CAL1
ON (Enters the calibration mode.)
CAL?
Response: CAL0 or CAL1
Calibration data
XINI
Initializes the calibration data.
(Not initializes the calibration data in the non-volatile
memory.)
XWR
Saves the calibration data to the non-volatile memory.
Calibration
execution
XVS
Selects the voltage source function calibration
XIS
Selects the current source function calibration.
XVLH
Selects the voltage limiter (High) calibration.
XVLL
Selects the voltage limiter (Low) calibration.
XILH
Selects the current limiter (High) calibration.
XILL
Selects the current limiter (Low) calibration.
XVM
Selects the voltage measurement function calibration.
XIM
Selects the current measurement function calibration.
Calibration range
XR-1
XR0
XR1
XR2
XR3
XR4
XR5
Sets the calibration range.
Calibration data
XDAT
Changes to the DMM data input mode.
XD
data: Inputs DMM read data.
XADJ
Changes to the calibration data fine adjustment mode.
XUP
Calibration data fine adjustment (UP)
XDN
Calibration data fine adjustment (DOWN)
XNXT
Moves to the next calibration.
*5:
Not initialized by the RINI command.
*6:
Not initialized by the RINI or *RST command.
Item
Command
Description
Default
Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS
Voltage range
Current range
XR-1
-
30
A
XR0
-
300
A
XR1
-
3 mA
XR2
-
30 mA
XR3
300 mV
300 mA
XR4
3 V
1 A
XR5
15 V
4 A