6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-35
System Comparator
calculation buzzer
BZ0
OFF
BZ1
ON (Comparator calculation result: HI)
BZ2
ON (Comparator calculation result: GO)
BZ3
ON (Comparator calculation result: LO)
BZ4
ON (Comparator calculation result: HI or LO)
BZ?
Response: BZ0
to BZ4
Limit detection
buzzer
UZ0
OFF
UZ1
ON
UZ?
Response: UZ0 or UZ1
Self test
*TST?
Executes the test and reads out the results.
Response: 0; Pass
1; Fail
TER?
The self-test result of each register is returned.
Response: a,b,c,d,e (a,b,c,d,e: 0 to 65535)
Error log
ERL?
Reads out error logs.
The number of errors and the error descriptions are all
cleared.
Response:
ddd,
ddd,
ddd,
ddd,
ddd
(+ is shown as a space.)
ERC?
Reads out the number of errors.
Response: ddd
000: No error
001 to 999: error count
(006 to 999 : to be overwritten)
Relay counter
RLY?
Reads out the relay counter data.
Response: ddddddddd (up to 999999999)
Interlock
OP0
STBY In signal input (IN)
OP1
OPR/STBY In signal input (IN)
OP2
InterLock In signal input (IN)
OP3
Operate Out signal output (OUT)
OP4
OPR/SUS In signal input (IN)
OP?
Response: OP0 to OP4
Item
Command
Description
Default
Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS