6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-32
Mea
sure
ment
Measurement
buffer memory
RN n[,adr]
n:
0... Releases recall execution status.
1... Sets recall execution status.
adr:
Recall data number (0 to 7999)
(The data number does not change if this set-
ting is omitted.)
Reading out the recall data by using the talker function
after the recall execution status is set performs the fol-
lowing operation.
•
Increments the recall data number after the data out-
put.
•
If data does not exist in the specified number, the
output becomes < EE + 8. 30 >
• Reading out does not erase the data in memory.
RN?
Response: RNn,adr
n : <d>
adr: <dddd>
RDN
adr1,adr2
Specifies the memory range to be read by RDT?.
(GPIB only)
adr1:
First recall data number (0 to 7999)
adr2:
Last recall data number (0 to 7999)
(0,0)
(0,0)
RDN?
Response: RDN adr1,adr2 (GPIB only)
adr1,adr2: <dddd>
RDT?
Reads out data from the specified memory range.
(GPIB only)
Response: Data is read out from the specified range in
the format described in Section 6.6, "Data Output For-
mat (Talker Format) using delimiters ",".
•
When data does not exist in the specified
number, <EE+8.30> is output.
•
Executing this command releases the recall
execution status.
(0)
SZ?
Reads out the number of stored data
Response: <dddd>
0
*6
RNM adr
adr:
Sets the maximum number of stored data (0
to 8000).
When the measurement memory is used and
the number of stored data in the buffer mem-
ory matches this value, bit 4 (ASM) of the
Device Event Status Register is set.
0
*6
RNM?
Reads out the maximum number of stored data.
Response: RNMdddd
*6:
Not initialized by the RINI or *RST command.
Item
Command
Description
Default
Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS