6240B DC Voltage Current Source/Monitor Operation Manual
5.2.10 Calculation Functions
5-55
5.2.10.2
Scaling Calculation
1. Calculation expression
The scaling calculation is defined using the following formula:
2. Operation
•
When the scaling calculation is ON, the MATH indicator goes ON.
•
The setting ranges of Constant A, Constant B, and Constant C are 0 to
999.999E+24 (however,
Constant A
0).
•
If the calculation result exceeds
9. 24, it is scaling-over and the error message
SCL
Over is displayed.
•
The scaling calculation is turned OFF by executing the *RST command.
•
Changing the measurement function does not turn OFF this calculation.
Scaling calculation =
Constant A
X-Constant B
Constant C
X: Measurement value