6240B DC Voltage Current Source/Monitor Operation Manual
4.2.6 MENU Key (Parameter Setting)
4-7
Random:
SWEEP VAL
Sweep Adr
Sets the start and stop addresses for random sweep.
Select the parameter by using the NULL (SEL) key.
Bias Value
Sets the bias value (source value before sweep start).
PSW Base
Sets the pulse sweep base value.
Linear 2:
SWEEP VAL
First Val
Sets the first value for 2-slope linear sweep.
Middle Val
Sets the middle value for 2-slope linear sweep.
Last Val
Sets the last value for 2-slope linear sweep.
Step1 Val
Sets the step 1 value between the first value and the middle value
for 2-slope linear sweep.
Step2 Val
Sets the step 2 value between the middle value and the last value
for 2-slope linear sweep.
Bias Value
Sets the bias value (source value before sweep start).
PSW Base
Sets the pulse sweep base value.
Figure 4-3 2-Slope Linear Sweep
TIME
Sets the time-related parameters for source and measurement
Hold Time
Sets the time from the sweep start to the step period start in the
sweep source mode.
Src Delay
Sets the delay time (Tds) from the start of the period time (Tp) to
the source start in the pulse source mode or sweep source mode.
Meas Delay
Sets the delay time (Td) from the measurement trigger to the mea-
surement start.
Start
Operate