Low-Level Disinfection
Prepare Enzol enzymatic detergent according to the manufacturer’s instructions. Spray
detergent solution liberally onto cuff and use a sterile brush to agitate the detergent so-
lution over entire cuff surface for five minutes. Rinse continuously with distilled water
for five minutes. To disinfect, first follow the cleaning steps above, then spray cuff with
10% bleach solution until saturated, agitate with a sterile brush over entire cuff surface
for five minutes. Rinse continuously with distilled water for five minutes. Wipe off excess
water with sterile cloth and allow cuff to air dry.
CAUTION:
Do not iron cuff.
CAUTION:
Do not heat or steam sterilize cuff.
Manometer Quality Control
A serial number and lot number are
automatically assigned to every aneroid
during manufacturing, ensuring every item
is "controlled."
The serial number can be located on the
faceplate of each aneroid (Figure 4).
The lot number is located on the box
end label (Figure 5).
Standards
ANSI/AAMI/ISO 81060-1:2007 • EN/ISO 81060-1:2012
Disposal
When your sphygmomanometer or any of its parts have reached their end of life,
please be sure to dispose of them in accordance with all regional and national
environmental regulations. Devices that have become contaminated should be
disposed of in accordance with all local ordinances and regulations.
Warranty
ADC
®
warrants its products against defects in materials and workmanship under
normal use and service as follows:
• Warranty service extends to the original retail purchaser only and commences
with the date of delivery.
•
See end panel label for package contents and to determine warranty duration.
Warranty duration is as follows:
Series
Manometer
Inflation System
760, 768, 769
20 Years
3 Years
780, 785, 790, 770, 775
10 Years
1 Year
What Is Covered:
Replacement of parts and labor.
What Is Not Covered:
Transportation charges. Damages caused by abuse, misuse, accident,
or negligence. Incidental, special, or consequential damages. Some states do not allow the
exclusion or limitation of incidental, special, or consequential damages, so this limitation
may not apply to you.
Implied Warranty:
Any implied warranty shall be limited in duration to the terms of this
warranty and in no case beyond the original selling price (except where prohibited by law).
This warranty gives you specific legal rights and you may have other rights which vary from
state to state.
To Obtain Warranty Service:
Send item(s) postage paid to: ADC, Attn: Repair Dept., 55
Commerce Dr., Hauppauge, NY 11788. Please include your name and address, phone no.,
proof of purchase, and a brief note explaining the problem.
For Australian Consumers
Our goods come with guarantees that cannot be excluded under the Australian Consumer
Law. You are entitled to a replacement or refund for a major failure and compensation for
any other reasonable foreseeable loss or damage. You are also entitled to have the goods
repaired or replaced if the goods fail to be of acceptable quality and the failure does not
amount to a major failure.
For our European Customers
On request, send to ADC by e-mail
(
), we can send to you this manual
on paper form within 7 calendar days at no additional cost to the user.
Our website,
https://www.adctoday.com
,
where these instructions for use are available
fulfills the requirements of personal data protection, according to Directive 95/46/EC and
GDPR on the protection of individuals with regard to the processing of personal data and
on the free movement of such data.
Any serious incident that has occurred in relation to this medical device should be reported
to ADC and the competent authority of the Member State in which the user and/or patient
is established.
Serial Number
Prosphyg
TM
Aneriods
Instructions for Use
Cuff Cleaning and Disinfecting
Use one or more of the following methods and allow to air dry:
• Wipe with mild detergent and water solution (1:9 solution). Rinse.
• Wipe with Enzol per manufacturer’s instructions. Rinse.
• Wipe with .5% bleach and water solution. Rinse.
• Wipe with 70% isopropyl alcohol.
• Launder with mild detergent in warm water, normal wash cycle.
Remove bladder first. Cuff is compatible with 5 wash cycles.
LATEX
(Figure 4)
FOR QUESTIONS, COMMENTS, OR SUGGESTIONS CALL TOLL FREE:
1-800-ADC-2670 or
visit www.adctoday.com/feedback
To register your product, visit us at
www.adctoday.com/support/warranty-registration
ADC
55 Commerce Drive
Hauppauge, NY 11788
U.S.A.
SC Cattus SRL
Aleea Carpati nr.53 Sc. B Ap. 64
Târgu-Mures, România, EU
Inspected, assembled
and packaged in the U.S.A.
tel: 631-273-9600
1-800-232-2670
fax: 631-273-9659
www.adctoday.com
email: [email protected]
1434
9360N-00 rev 15.qxp_Layout 1 6/13/22 1:37 PM Page 1