Low-Level Disinfection

                                   

Prepare Enzol enzymatic detergent according to the manufacturer’s instructions. Spray 
detergent solution liberally onto cuff and use a sterile brush to agitate the detergent so-
lution over entire cuff surface for five minutes.  Rinse continuously with distilled water 
for five minutes.  To disinfect, first follow the cleaning steps above, then spray cuff with 
10% bleach solution until saturated, agitate with a sterile brush over entire cuff surface 
for five minutes.  Rinse continuously with distilled water for five minutes.  Wipe off excess 
water with sterile cloth and allow cuff to air dry.

 

 

CAUTION:

 Do not iron cuff.

 

CAUTION:

 Do not heat or steam sterilize cuff.

 

 

 

Manometer Quality Control

 

A serial number and lot number are                   

 

automatically assigned to every aneroid            
during manufacturing, ensuring every item       
is "controlled."  

 

The serial number can be located on the           
faceplate of each aneroid (Figure 4). 

 

The lot number is located on the box              
end label (Figure 5).  

 

 

 

 

 

 

 

 

  

 

 

 

 

 

 

 

Standards

 

    ANSI/AAMI/ISO 81060-1:2007 • EN/ISO 81060-1:2012

 

 

 

  
Disposal 

When your sphygmomanometer or any of its parts have reached their end of life, 
please be sure to dispose of them in accordance with all regional and national 
environmental regulations. Devices that have become contaminated should be 
disposed of in accordance with all local ordinances and regulations.

 

Warranty

 

ADC

®

 warrants its products against defects in materials and workmanship under 

normal use and service as follows:

 

 

•  Warranty service extends to the original retail purchaser only and commences 

    with the date of delivery. 

•  

See end panel label for package contents and to determine warranty duration.

 

 

 

Warranty duration is as follows:

 

    

Series

Manometer

Inflation System

 

760, 768, 769

20 Years

3 Years 

780, 785, 790, 770, 775

10 Years

1 Year 

 

What Is Covered:

 Replacement of parts and labor. 

What Is Not Covered:

 Transportation charges. Damages caused by abuse, misuse, accident, 

or negligence. Incidental, special, or consequential damages. Some states do not allow the 
exclusion or limitation of incidental, special, or consequential damages, so this limitation 
may not apply to you. 

Implied Warranty:

 Any implied warranty shall be limited in duration to the terms of this 

warranty and in no case beyond the original selling price (except where prohibited by law). 
This warranty gives you specific legal rights and you may have other rights which vary from 
state to state. 

To Obtain Warranty Service:

 Send item(s) postage paid to: ADC, Attn: Repair Dept., 55 

Commerce Dr., Hauppauge, NY 11788. Please include your name and address, phone no., 
proof of purchase, and a brief note explaining the problem. 

 

For Australian Consumers

 

Our goods come with guarantees that cannot be excluded under the Australian Consumer 
Law. You are entitled to a replacement or refund for a major failure and compensation for 
any other reasonable foreseeable loss or damage. You are also  entitled to have the goods 
repaired or replaced if the goods fail to be of acceptable quality and the failure does not 
amount to a major failure. 

 

For our European Customers

 

On request, send to ADC by e-mail 

(

[email protected]

), we can send to you this manual 

on paper form within 7 calendar days at no additional cost to the user. 

 

 

Our website, 

https://www.adctoday.com

where these instructions for use are available 

fulfills the requirements of personal data protection, according to Directive 95/46/EC and 
GDPR on the protection of individuals with regard to the processing of personal data and 
on the free movement of such data.  

 
Any serious incident that has occurred in relation to this medical device should be reported 
to ADC and the competent authority of the Member State in which the user and/or patient 
is established. 

 

Serial Number

Prosphyg

TM

 Aneriods 

Instructions for Use

Cuff Cleaning and Disinfecting

 

Use one or more of the following methods and allow to air dry: 

•   Wipe with mild detergent and water solution (1:9 solution).  Rinse. 

•   Wipe with Enzol per manufacturer’s instructions.  Rinse. 

•   Wipe with .5% bleach and water solution.  Rinse. 

•   Wipe with 70% isopropyl alcohol. 

•   Launder with mild detergent in warm water, normal wash cycle. 

      Remove bladder first. Cuff is compatible with 5 wash cycles. 

LATEX

(Figure 4)

 

FOR QUESTIONS, COMMENTS, OR SUGGESTIONS CALL TOLL FREE: 

1-800-ADC-2670 or 

visit www.adctoday.com/feedback

 

To register your product, visit us at  

www.adctoday.com/support/warranty-registration

              ADC 
              55 Commerce Drive 
              Hauppauge, NY 11788 
              U.S.A. 
 
              

SC Cattus SRL 

             Aleea Carpati nr.53 Sc. B Ap. 64 
             Târgu-Mures, România, EU

Inspected, assembled 

and packaged in the U.S.A. 

 

tel: 631-273-9600 

1-800-232-2670 

fax: 631-273-9659 

 

www.adctoday.com 

email: [email protected]

1434

9360N-00 rev 15.qxp_Layout 1  6/13/22  1:37 PM  Page 1

Reviews: