ADView 2 User Manual | 30
93-9005-00
rev 4
September 1, 2016
The tested device meets the stated accuracy specification of RMS Error = 2.00% for the range 70% to 100% SaO
2
. The
accuracy is not specifed below 70% SaO
2
. In addition, here are detailed measured A
RMS
values over the specified
discreet ranges:
Range
A
RMS
90-100%
0.60%
80-90%
0.54%
70-80%
0.67%
Special Notes for ChipOx SpO
2
module:
NOTE: Use only SpO2 sensors from ADC approved for use on the ADView 2 ChipOx SpO2 module.
NOTE: A functional tester cannot be used to assess the accuracy of a ChipOx pulse oximeter sensor or pulse
oximeter. Using FLUKE Biomedical Index 2 SpO
2
simulator, the device can display the entire range of SpO
2
values.
ChipOx Sensors:
Before using, carefully read the ChipOx sensor directions for use, which includes cleaning and disinfection
instructions.
ChipOx sensors are provided non-sterile.
ChipOx Graphical Plot:
Below is a graphical plot of discreet A
RMS
values measured with ChipOx Oximetry in a clinical study using ChipOx
pulse oximetry sensors: