background image

 

CLEARPASS 

Full-Body Security Screening System 

 

 

 

OPERATION  
MANUAL

 

Identification 

ADN190.00.00.000 OM2

 

Revision  

1

 

Date 

02.2022

 

Page 

26 of 52

 

SYSTEM DESCRIPTION 

SELECTION 3 

Summary of Contents for CLEARPASS

Page 1: ...2 ADANI Systems Inc 13631 Poplar Circle Conroe Texas 77304 Technical Support Tel 844 989 6789 support adanisystems com Sales Inquiries Tel 936 588 2064 sales adanisystems com FULL BODY SECURITY SCREEN...

Page 2: ...service specialists and contains basic information about the System instructions on safety measures and compli ance with the requirements of regulatory documents design description information on prep...

Page 3: ...LITIES 14 2 3 RADIATION SAFETY 15 2 4 ELECTRICAL SAFETY 23 2 5 FIRE SAFETY 24 2 6 MECHANICAL SAFETY 24 2 7 INFORMATION SAFETY 24 2 8 EMERGENCY ACTIONS 25 3 SYSTEM DESCRIPTION 27 3 1 FRAMEWORK 28 3 2 P...

Page 4: ...2 2022 Page 4 of 52 5 3 EMERGENCY SHUTDOWN 44 6 MAINTENANCE 46 6 1 GENERAL 46 6 2 SAFETY INSTRUCTIONS WHEN PERFORMING THE MAINTENANCE 46 6 3 MAINTENANCE PROCEDURE 46 6 4 PERIODIC MAINTENANCE 48 6 5 DI...

Page 5: ...symbols are affixed to the exterior of the System Refer to the accompanying documents Radiation hazard Caution Dangerous Voltage INDICATES HAZARDOUS SITUATIONS WHICH IF NOT AVOIDED WILL RESULT IN DEAT...

Page 6: ...alified as well as able to undertake all required safety precautions The manufacturer his her authorized agents and representatives are not liable for injury or damage that may result from System impr...

Page 7: ...CLEARPASS Full Body Security Screening System OPERATION MANUAL Identification ADN190 00 00 000 OM2 Revision 1 Date 02 2022 Page 7 of 52 GENERAL SELECTION 1...

Page 8: ...raphic images of a person in full height The general view of the System without Operator s Acquisition Analysis workstation fur ther Operator s AWS is shown in Figure 1 1 Figure 1 1 System General Vie...

Page 9: ...in a fixed position on the X ray detector with his her back to the detector as close as possible see Figure 5 1 2 When starting the scanning process the X ray generator is turned on and the rotary fr...

Page 10: ...rs in case of use data acquisition from detector image creation and saving as a file digital processing of the images printing of the digital X ray images The software provides a variety of tools to p...

Page 11: ...ody swallowed or hidden in body cavities at a single inspection high throughput no personal search required elimination of the influence of uninformative scattered radiation on image quality Modern di...

Page 12: ...ATION MANUAL Identification ADN190 00 00 000 OM2 Revision 1 Date 02 2022 Page 12 of 52 The location of the symbols and identification plate on the System is shown in Figure 1 4 Figure 1 4 Location of...

Page 13: ...CLEARPASS Full Body Security Screening System OPERATION MANUAL Identification ADN190 00 00 000 OM2 Revision 1 Date 02 2022 Page 13 of 52 SAFETY SELECTION 2...

Page 14: ...FOLLOW THE INSTRUCTIONS GIVEN IN THIS MANUAL FOR A LONG TERM AND SAFE OPERATION OF THE SYSTEM READ THOROUGHLY THIS MANUAL PRIOR TO USING THE SYSTEM AND KEEP IT AT THE OPERATOR S AWS THE REQUIREMENTS...

Page 15: ...using X rays ANSI HPS 43 17 2009 Directive 2013 59 EURATOM Recommendations of the International Commission on Radiation Protection ICRP 2007 Recommendations Federal Food Drug and Cosmetic Act FFDCA Ch...

Page 16: ...vement In case of deviation from the conditions and parameters of the mechanism and electronic devices of the System set by the manufacturer as well as with their failure the software disa bles the X...

Page 17: ...poses So the System is designed in compliance with the requirements of ANSI HPS No 43 17 2009 When operating the System a high level of radiation safety is ensured by the lowest possible doses and the...

Page 18: ...The reference effective dose is in the range of 0 25 Sv 4 0 Sv 25 rem 0 4 mrem per screening withing range 3 5 sec of scanning time The reference effective dose received by individuals from one facili...

Page 19: ...dose delivered to a person for single inspection in General use radiation dose delivered to a person for single inspection in Limited use 0 25 2 For medical researches the maximum permissible doses a...

Page 20: ...ng equipment operation In addition it is neces sary to determine the size of the sanitary protection zone in order to ensure the equivalent level of an ambivalent dose of no more than 1 mSv per year A...

Page 21: ...maximum throughput at maximum operation conditions Figure 2 2 Sanitary protection zone To place the System in accordance with the requirements of national radiation protection laws the user independe...

Page 22: ...pened allows intrusion into the radiation field 10 Interlocking devices terminate the primary beam output in the event of any System faults that could result in abnormal or unintended radiation emissi...

Page 23: ...zed in case of a single fault 2 4 1 Electromagnetic compatibility The System meets the requirements for electromagnetic compatibility requirements es tablished in IEC 61326 1 2013 EN 55011 2016 and El...

Page 24: ...the System is operated the following dangerous situations can occur risk of stumbling risk of getting caught on the wires 2 7 INFORMATION SAFETY BE CAREFUL WHEN CONTROLLING THE MOTION OF THE SYSTEM M...

Page 25: ...or the control console see Figure 4 2 and if necessary provide first aid to the victim Next disconnect the System and the operator s computer from the mains supply and then contact the organization t...

Page 26: ...CLEARPASS Full Body Security Screening System OPERATION MANUAL Identification ADN190 00 00 000 OM2 Revision 1 Date 02 2022 Page 26 of 52 SYSTEM DESCRIPTION SELECTION 3...

Page 27: ...ystem as well as the location and purpose of its main functional parts The major functional parts of the System are as follows see Figure 3 1 Framework Portal External covers Signal Lamps X ray tube m...

Page 28: ...ting the System on working place Side sections position 2 and 4 see Figure 3 2 are for structural rigidity and used to mount the System external covers The top cover position 3 see Figure 3 2 is cover...

Page 29: ...and saving images to the CONPASS database 3 3 EXTERNAL COVERS The external covers see Figure 3 4 of the System restrict access to the internal com ponents and units and also protect them from mechanic...

Page 30: ...em for Operation Figure 3 4 Signal Lamp Assemblies The System external covers are equipped with a panel to control cables and power sup ply cables output 3 4 X RAY TUBE MODULE The X ray tube module in...

Page 31: ...of 52 1 Rotary mechanism 2 X ray Detector 3 Rotary frame 4 Collimator output slit 5 X ray generator with Beam limiting unit 6 Electrics Figure 3 5 3 4 1 Rotary frame The rotary frame see Figure 3 6 i...

Page 32: ...52 Figure 3 6 3 4 2 X ray detector The X ray detector see Figure 3 7 is a multipurpose photodiode array element with scintillation coating The current of the photodiodes is proportional to the total...

Page 33: ...nts photodiode scintillator combination is charac terized by the number of elements pixels of a digital image During the scanning process the output signal of each element is successively read digital...

Page 34: ...person scanning 1 X ray Generator 2 Generator Control Unit 3 X ray Generator Mounting Mechanism 4 Beam Limiting Unit Figure 3 8 X ray Generator The X ray generators are inherently protected disabling...

Page 35: ...tion control system The inbuilt X ray filter ensures the maximum quality of the image with the minimum dose rate for the inspected individual The filtering cuts off the soft part of X radiation that c...

Page 36: ...nsmitted through the coupling using a belt position 4 see Figure 3 10 to the lower rotary unit position 3 see Figure 3 10 The lower unit moving along the guide on the framework base plate moves the ro...

Page 37: ...ry Mechanism Figure 3 10 Rotary Mechanism 3 5 OPERATOR S ACQUISITION ANALYSIS WORKSTATION OPERA TOR S AWS Operator s AWS is a software hardware complex to control the System operation to acquire the f...

Page 38: ...stem obtain data from the X ray dosimeters if available acquire view graphically process and archive X ray images Operator s AWS can be supplied with or without Operator s stand 3 6 VIDEO SURVEILLANCE...

Page 39: ...CLEARPASS Full Body Security Screening System OPERATION MANUAL Identification ADN190 00 00 000 OM2 Revision 1 Date 02 2022 Page 39 of 52 SETUP PROCEDURES SELECTION 4...

Page 40: ...the workstation stand computer and entering the operating system the operator s program will start automatically 3 Warm up the X ray generator in case it has not been used for a long time and the WAR...

Page 41: ...icator turns green then 1 Key 2 LED Indicator Push Button 3 Emergency Stop Button Figure 4 2 Control Box 3 Press and hold the button on the UPS which is part of the Operator s AWS until you hear an au...

Page 42: ...CLEARPASS Full Body Security Screening System OPERATION MANUAL Identification ADN190 00 00 000 OM2 Revision 1 Date 02 2022 Page 42 of 52 OPERATION SELECTION 5...

Page 43: ...n the moveable platform or the conveyor belt as shown in Figure 5 1 Figure 5 1 Positioning the Inspected Individual during Inspection 3 Perform the scanning as stipulated in the Operator s Manual 4 Re...

Page 44: ...er parts of the system There are 2 buttons on it window Press RESUME if you want time to fix the problem Press SHUT DOWN if it s emergency problem and you need to turn off the system ex Emergency butt...

Page 45: ...CLEARPASS Full Body Security Screening System OPERATION MANUAL Identification ADN190 00 00 000 OM2 Revision 1 Date 02 2022 Page 45 of 52 MAINTENANCE SELECTION 6...

Page 46: ...iodic M2 and M3 of the System are performed by the service specialists of the man ufacturer or by the service organization authorized by the manufacturer 6 2 SAFETY INSTRUCTIONS WHEN PERFORMING THE MA...

Page 47: ...se 0 5 soap solution of the detergent Check the functionability of emergency stop buttons on the System see Figure 2 1 and on the control console see Figure 4 1 as follows Switch on the System as stip...

Page 48: ...tions 2 Check insulation of electric wires monitor ground cir cuit resistance 3 Remove housings inspect PCBs of system elec tronic modules wash PCB connector pins in alcohol and install the housings b...

Page 49: ...he components of the System Remove lead protection elements electronic units cables and wires Remove accumu lators from the UPS Remove electronic boards from the housings of electronic equipment syste...

Page 50: ...CLEARPASS Full Body Security Screening System OPERATION MANUAL Identification ADN190 00 00 000 OM2 Revision 1 Date 02 2022 Page 50 of 52 TECHNICAL SPECIFICATION SELECTION 7...

Page 51: ...th 87 4 2220 mm width 63 1600 mm height 94 5 2400 mm Weight max 1874 lbs 850 kg Supply ratings Current type phase number Voltage V Frequency Hz Power consumption kV A max single phase AC protective ea...

Page 52: ...plex Operator s Manual 7 3 ENVIRONMENTAL REQUIREMENTS Operation Air temperature limits Max relative humidity at 68 F 20 0 to 45 32 to 113 F up to 90 Transportation Air temperature F Max relative humid...

Reviews: