MODEL: F171A (

230V)

Summary of Contents for F171A

Page 1: ...READ MANUAL CAREFULLY FOR PROPER USE AND OPTIMAL OPERATION 14841 Dallas Parkway Suite 500 Dallas TX 75254 ActivePure Medical Healthcare Inspired Science Powered MODEL F171A 230V ...

Page 2: ...e outdoors or on wet surfaces DO NOT look directly at the UV light inside unit when operating eye damage may occur DO NOT allow device to be used as a toy do not permit children to operate device DO NOT use with damaged cord or plug if device is not working as intended has been dropped damaged left outdoors or is wet contact Customer Service for support at 1 800 572 6241 NEVER pull or carry by cor...

Page 3: ...ter Replacement Instructions 9 ActivePure Cell Replacement Instructions 11 Replacement Parts 12 FAQs 12 Warranty Information 13 Device IEC 60601 1 2 2014 EMC Information 14 The ActivePure Medical Guardian model F171A is a device intended for medical purposes that is used for the reduction of staphylococcus epidermidis and erwinia herbicola bacteria MS2 and Phi X174 viruses and aspergillum niger fu...

Page 4: ...rnal fan to draw air through the device as shown in the general device design Figure 1 Lab test results demonstrated the ActivePure Medical Guardian model F171A device using ActivePure to expose Staphylococcus epidermis and MS2 bacteriophage to hydroxyl radicals and super ions formed as the result of UV exposure on the photo catalyst has the ability to destroy the tested bio organisms on the filte...

Page 5: ...ly charged particles which flow out from a series of carbon brushes Carbon to remove odors Hydrophobic polypropylene pleated media filter designed to reduce airborne contaminants Powerful UVGI bulbs Photo catalyst module using ActivePure Technology 4 Speed fan draws airborne contaminants through the device All steel and aluminum construction for durability and long life Automatic ambient light sen...

Page 6: ...x 21 0 D 673mm H x 292mm W x 533mm D Weight 48 lbs 21 8 kg Power Cord 6 ft 1 8 m ActivePure Medical LLC reserves the right to change or modify any specifications without notice Consult Activepure com for the latest device information or consult Customer Service at 1 800 572 6241 Life 3 years minimum operational life 1 Place the ActivePure Medical Guardian in a location that allows air to move free...

Page 7: ...eople or equipment within the space Rates of ventilation The device is intended to result in a 4 6 log reduction of airborne microorganisms from initial concentrations in 60 minutes as shown in Table 1 The ActivePure Medical Guardian model F171A is not a sterile device and is not intended to create a sterile environment Organism Type Organism Name Initial Load of Microorganisms Avg Log Reduction a...

Page 8: ... safety shutdown feature to ensure the filter and or cell replacement is performed at the appropriate times When the filter and or cell replacement time is reached the appropriate LED indicator will illuminate However if the replacement indicator s are ignored or if the one or both LED s fail to illuminate by design the device will shut down 3 days of operation after reaching the replacement time ...

Page 9: ... instruction for use cleaning disinfecting filter replacement sheet Step 4 Before removing and handling the used filter put on the disposable mask and gloves Step 5 Slide remove the used filter from the device placing it inside the large plastic bag provided for disposal then set it aside Step 6 Use Metrex CaviWipes to clean the device These are NOT provided with the device or replacement filter N...

Page 10: ...e rub the wipe over the surface until soil is no longer visible on this part of the device If soil is heavy or is still visible repeat step with a new CaviWipe until soil is no longer visible Step 15 Allow the device to dry no visible wetting remains No rinsing is required With the device cleaned it is ready for disinfecting DISINFECTING Disinfectants are typically provided as a concentrate to be ...

Page 11: ... reduction of staphylococcus epidermidis and erwinia herbicola bacteria MS2 and Phi X174 viruses and aspergillum niger fungal spores and bacillus globigii bacterial spores from the air in a temperature controlled professional healthcare environment of 70 71 F 40 45 RH The following instructions will permit the device ActivePure Cell to be safely replaced without creating safety concerns in the pat...

Page 12: ...and decide not to change the filter or cell A The device has a safety feature to prevent the continued use of the device 3 days after it is time to change the filter or cell Q How do I reprocess the device A Follow the instruction inside the user manual on the website or what comes with a replacement filter These instruction details the cleaning and disinfecting procedure Q What if I have question...

Page 13: ...only Warranties are voided if a device is purchased through unauthorized channels this includes websites that are not authorized to use ActivePure Medical s trademarked names images and logos as well as Internet auction sites e g ebay and Craigslist The only approved Internet presence for ActivePure Medical LLC products is Activepure com To confirm warranty coverage prior to purchasing a device co...

Page 14: ... the EMC requirement of IEC 60601 1 2 and is suitable for use in a professional healthcare environment This device is not effected by RFID wireless networks 2 ways radios paging systems etc as the device does not transmit or receive RF signals or create EMF beyond the required limits Do not use device with an extension cord as unintended EMF may be created This device is grounded to prevent electr...