7.4.1 I-Surge mode factory settings
Programme
Contra-angle ratio
Speed at end of tool
Torque at end of tool
Irrigation
P1
20:1
1200 rpm
80 N.cm
80 ml/min
P2
20:1
800 rpm
80 N.cm
100 ml/min
P3
20:1
15 rpm
20 N.cm
100 ml/min
P4
20:1
30 rpm
20 N.cm
0
7.4.2 Piezotome mode factory settings
Programme
Power
Irrigation
Very powerful
D1
3
60 ml/min
Powerful
D2
3
60 ml/min
Average power
D3
3
60 ml/min
Low power
D4
3
60 ml/min
7.4.3 Newtron mode factory settings
Programme
Power
Irrigation
Main functions
Green
Soft
P = 5
15 ml/min
Periodontics
Yellow
Medium
P = 5
15 ml/min
Endodontics
Blue
High
P = 5
15 ml/min
Scaling
Orange
Boost
P = 5
15 ml/min
Loosening
7.5 Personalising a contra-angle or handpiece
1. Switch on the medical device.
2. Press
.
3. Press
.
4. Enter the handpiece contra-angle ratio as specified by the manufacturer, using the keypad
5. Rpm and Ncm values are recalculated automatically.
6. Press
to save the values.