Pictograms
for the I-Surge
mode
Meaning
Select programme P4 - Screwing
Clockwise rotation
Anti-clockwise rotation
I-Surge mode programme settings
Programme Contra-angle
ratio
Speed
at end of tool
Torque
at end of tool
Irrigation
Function
P1
20:1
1200 rpm
80 N.cm
80 ml/min
Implant site marking
P2
20:1
800 rpm
80 N.cm
100 ml/min
Pilot drilling
P3
20:1
15 rpm
20 N.cm
100 ml/min
Boring/drilling
P4
20:1
30 rpm
20 N.cm
0
Screwing
5.2 Piezotome
®
M+ mode
Before using an M+ tip, read the M+ tip user manuals [J87601] and [J87611] and the Quick Reference
guide for M+ tips [J87621].
Pictograms
for the Piezotome
mode
Meaning
Power
Select programme D1 - Osteotomy, osteoplasty
Very powerful
Select programme D2 - Osteotomy, osteoplasty
Powerful
Select programme D3 - Osteotomy, osteoplasty
Average power
Select programme D4 - Separation of soft tissue
Low power
Used power per programme - press the + and - pictograms to adjust the value
Adjust the power
Programme
Power
Irrigation
D1
3
60 ml/min
D2
3
60 ml/min
D3
3
60 ml/min
D4
3
60 ml/min
Fine setting adjusts the power level for each programme between 1 and 5. The power value is a percentage of
the maximum power at D1-5.
Programme
D1
Fine setting level
1
2
3
4
5
Power value
82%
86%
90%
95%
100%
Frequency modulation
60 Hz
User Manual • ImplantCenter™ M+ • J27251 • V1 • (13) • 09/2013 • NO25EN010A - Page
15
/
42
Summary of Contents for ImplantCenter M+
Page 1: ...User Manual ImplantCenter M...
Page 2: ......
Page 5: ...15 Glossary 41...
Page 6: ......
Page 8: ......
Page 26: ...User Manual ImplantCenter M J27251 V1 13 09 2013 NO25EN010A Page 24 42...
Page 28: ...User Manual ImplantCenter M J27251 V1 13 09 2013 NO25EN010A Page 26 42...
Page 40: ...User Manual ImplantCenter M J27251 V1 13 09 2013 NO25EN010A Page 38 42...